N O T E
9 Click Run in the Autotune dialog box.
Agilent 7500 ICP-MS ChemStation Operator's Manual
Tuning
box.
Autotune tuning parameter settings can also be saved as an Autotune Target File. For
the details, see the chapter "Creating and Using a Autotune Target File".
The selected items in tune will be executed in the Autotune Mode. During autotune,
Autotune Situation dialog box will be displayed.
The Autotune Situation Dialog Box
The status of each tuning item is displayed in color. The colors have the following
meaning:
• Gray
This tuning item was not selected in the Autotune dialog box.
• White
This item was selected in the Autotune dialog box, but has not been executed yet.
• Yellow
Auto tuning for this item is currently being executed. The progress of tuning is
being displayed in the status bar.
• Red
Autotune was unable to tune this item within set targets. In some cases Autotune
may executed this item again after the optimization of other parameters.
• Blue
Tuning is complete (shows that it has been executed as per target).
The Status Bar shows a graphical representation of the progress of each item as
4-109