Agilent 7500 ICP-MS ChemStation Operator's Manual
Performing a Quantitative Analysis
The three most recently loaded data files are listed as numbered menu items, with
the first entry corresponding to the most recently loaded data file. Selecting any one
of these numbered menu items will cause that file to be loaded into data analysis.
2 Select FullQuant>>Generate Report.
The FullQuant Report Option dialog will appear.
3 Select the report style.
The following report styles are available:
• Summary
Gives the counts per second (CPS) or counts (for Time Resolved Analysis), con-
centration, and relative standard deviation (RSD) for each mass in the sample.
• Detailed (Text Only)
Gives the same information as the summary report, as well as, the standard devi-
ation (SD). The format is slightly different but the SD and RSD for the counts
per second (CPS) or counts (for Time Resolved Analysis) and the concentration
for each mass are still displayed.
• Detailed
Includes the same information as the Detailed (Text Only) report, but which also
includes a spectrum of the sample or a time chart of each element at the bottom
of the report. This report can only be sent to the printer.
4 Select the destination for the report.
You can click one or more of the following check boxes:
• Screen
Generates the report to the ChemStation screen.
• Printer
Generates the report to the ChemStation printer. You must select this destination
for a detailed report.
• File
Generates the report to the file you indicated in the text box.
5 Click OK.
The report will be generated to the selected destination in the selected style.