Agilent 7500 ICP-MS ChemStation Operator's Manual
Appendix
5 Select Tune >> Dead time calibration..
Select Element for EM Dead time Calibration dialog box is displayed.
6 Select Er[164/166] then click OK..
Select Element for EM Dead time Calibration dialog box
7 Check sensitivity for 50ppb Er
Select Check .. the sensitivity must be above 50,000 counts. Select OK then Run ..
A-32