Agilent 7500 ICP-MS ChemStation Operator's Manual
Tuning
When selected, autotune automatically optimizes the torch vertical and horizontal
position.
Adjusted parameters:
Torch-H, Torch-V
• Lens/Plasma
When selected, autotune automatically optimizes the ion lens parameters in
accordance with the target values.
Adjusted parameters:
Extraction Lenses, Einzel Lenses, Omega Lenses, QP Focus, Plate Bias,
RF Power, Smpl Depth, Carrier Gas, Makeup Gas
• Tuning Report
Check this box to generate a tuning report after autotune is completed.
If the Record log for each tuning report check box is checked on the Detail Set-
ting dialog box, opened from the ICP-MS Configuration dialog box, the tuning
results will be automatically recorded in the maintenance log.
7 Enter the vial number of the tuning solution in the ALS Tuning Solution Vial
text box.
When the autosampler is set in the Configuration task, this text box appears. When
"0" is entered, the tuning starts without moving the sample probe of the autosam-
pler.
8 Set the target values for tuning, if necessary.
When you click Target Setting, the Set Autotune Targets dialog box is displayed.
The Set Autotune Targets dialog box allows the selection of target values used by
autotune.
4-105