Agilent 7500 ICP-MS ChemStation Operator's Manual
Tuning
When you click the Range Setting, the Parameter Range appears at the right side of
the Set Autotune Target dialog box. In autotune it is possible to set a range for each
variable tuning parameter value.
The Set Autotune Targets Dialog Box
Each parameter has been assigned into one of 4 categories: Plasma Parameters,
Lens parameters, Mass Parameters, and EM Parameters. There are variable ranges
or fixed values parameters which can be set for the categories.
In the Plasma Parameters and Lens Parameters, it is possible to select the
variable/fixed parameter (the variable range in the case of selecting variable and the
fixed value in the case of selecting fixed). When the present value is shifted from the
variable range, the present value will be displayed in red.
It is possible to carry out tuning effectively by providing an appropriate fixed value.
If you click Reset to Default, The factory default values are reset.
On completion of the target value setting, click OK to return to the Autotune dialog