ASX500 Configuration Dialog Box
Agilent 7500 ICP-MS ChemStation Operator's Manual
Configuration
Only one LC/GC ChemStation can synchronize with the ICP-MS ChemStation, and
you can control only a single instrument from the LC/GC ChemStation.
To use the LC/GC ChemStation, you must enable Remote Start. Selecting the Link
to Agilent LC/GC ChemStation checkbox prevents selection of Don't Use in Remote
Start.
• ISIS application
When ISIS is selected as Type, an appropriate ISIS application, which are either
standard or use defined application, should be selected in the pull-down menu.
• Autosampler
If you want to use the autosampler (ASX500, ASX100, or I-AS), select the
autosampler type in the pull-down menu. If an Autosampler is not being used,
select None.
If an Autosampler type is selected, set the communication port and the sample
rack size for the autosampler.
The ASX500, ASX100, or I-AS configuration dialog box appears when you click
Setup.
ASX100 Configuration Dialog Box
I-AS Configuration Dialog Box