Agilent 7500 ICP-MS ChemStation Operator's Manual
Tuning

Autotune

Autotune automates the optimization of various tunable lens and voltage settings to
optimize the instrument. In Autotune it is possible to change the targeted sensitivity
along with the range for each tuning parameter.
RF matching, Optional Gas (in case the optional gas line is installed), PeriPump, S/C
temp, and QP Bias are not optimized with Autotune. Optimize these parameters
manually before Autotune.
To adjust the RF Matching, use of the automatic adjustment function is
recommended. To use the automatic adjustment function, select
Tune>>RF Matching on the Tuning window, and then follow the displayed dialog
box.
During Autotune, do not use other software functions.
If you use an autosampler, set Autosampler in the Configuration task. For more
information about the configuration task, see Chapter 2, "Configuration". When you
set the configuration for an autosampler, you will need to turn on the autosampler
before running Autotune.
4-102