Agilent 7500 ICP-MS ChemStation Operator's Manual
Running a Sample Analysis

Stopping a Run

Stopping a Run from the Real-Time Display Window

To stop a run from the Real-Time display before the run time elapses, click the Stop
Run push button. The Agilent 7500 will stop in the middle of the current repetition.
When you manually stop the run, data acquired for the current repetition is not saved.
You can start the run again by clicking the Restart push button. ChemStation starts
acquisition on the current sample using the sample information already provided. If
you must change the sample information before running the sample, click Acquire
Data>>Acquire Data in the ICP-MS Acquisition window, type the new sample
information in the dialog box that appears, then click the Acquire push button.

Stopping a Run by Stopping a Method

If you want to stop the method you are currently running, click the Abort push but-
ton from the Run Method status window. A dialog box will pop up asking if you are
sure you want to abort. Click Yes, and the entire method becomes inactive.