Agilent 7500 ICP-MS ChemStation Operator's Manual
Creating a Method
2 Select the Calibration Save Options by clicking on the boxes for each option.
There are 2 options available on the Calibration Save Options dialog box:
• Click the Alert when Calib is Overwritten box if you want the ChemStation to
alert you with a message when an existing calibration will be overwritten.
• Click the Verify QC Method box if you want ChemStation to verify the QC
method.
3 After selecting the options, click OK.
If you have selected the Verify QC Method option, ChemStation will verify the QC
method.
After verifying the QC method, the Save Calibration As dialog box.
4 Give your new Calibration a name and then click OK.