Tuning the ShieldTorch (Cool plasma)

Ca (40 AMU) can be analyzed with the cool plasma conditions not the normal
plasma conditions. To do that, delete the 40 AMU in the Prohibited Masses of the
Configuration task.
When you return to the normal plasma conditions from the cool plasma condition,
you must reset the 40 AMU in the Prohibited Masses because detector gets damage.

Tuning solution

Li, Y, Ce, and Tl solution at 1 ppb: for tuning the mass resolution/axis
Co solution at 1 or 10 ppb: for tuning the sensitivity
In the cool plasma condition, Y cannot be used because it generates the oxide ion
(YO), which decreases the sensitivity of 89amu(Y).

Tuning procedure

To use the ShieldTorch in the cool plasma condition, perform the following steps: