Startup, Shutdown and Status

Startup, Shutdown and Status
The Agilent Technologies, Agilent 7500 is designed so that it is easy to start up and
shut down. The instrument has three states of operation that are displayed in the
Agilent 7500 ChemStation instrument control software:
• Analysis
During tuning and sample analysis, the Agilent 7500 is in Analysis mode.
• Standby
At the end of the working day, place the Agilent 7500 in Standby mode.
In this mode both the turbo pumps and the backing rotary pumps remain on, but
the quadrupole, detector, and RF (plasma) are switched off.
• Shutdown
To maintain or move the Agilent 7500, put it into Shutdown mode. In an emer-
gency, you can flip the main power switch to stop the instrument.
This chapter explains how to start and shut down the Agilent 7500, as well as, check
instrument status.