E Diagnostics quick reference

AM Dispense task parameters

AM Dispense task parameters

About this topic
Parameter description
276
This topic describes the task parameters for the AM Dispense task, which is
available only for the Bravo 96AM Head.
The AM Dispense task is designed for dispensing fluids through AssayMAP Bravo
cartridges that are mounted on the Bravo 96AM Head. In addition to the parameters
for the Dispense task, the AM Dispense task allows you to override the flow rate from
liquid class.
To run the AM Dispense task from Diagnostics, see
Diagnostics" on page
100. To use the task in a protocol, see the
Control User
Guide.
The AM Dispense task has the following parameters.
Parameter
Empty tips
Volume (µL)
Blowout volume (µL)
Liquid class
Override dispense flow rate
from liquid class
"Performing a task using Bravo
Description
The option to empty all liquid from the probes,
cartridges, or tips instead of using the dispense
volume specification.
The volume of liquid to be dispensed from each
probe, cartridge, or tip.
Specifies the volume of air to dispense after the
main volume has been dispensed while the
probes, cartridges, or tips are still in the wells.
Typically, the blowout volume is the same as the
pre-aspirate volume.
Note: Blowout occurs only in the last quadrant
dispensed for a given AM Dispense task.
The pipetting speed and accuracy.
IMPORTANT
To ensure consistent pipetting,
always select a liquid class for liquid-handling
tasks.
The option to override the dispense velocity in the
selected liquid class. Selecting this option enables
you to specify a value for the dispense flow rate
without changing the liquid class.