E Diagnostics quick reference

Dispense task parameters

Dispense task parameters

About this topic
Parameter descriptions
250
The Dispense task dispenses liquid into a microplate or reservoir. This topic
describes the task parameters.
To run the Dispense task, see
page
100.
The Dispense task has the following parameters.
Parameter
Empty tips
Volume (µL)
Blowout volume (µL)
Liquid class
Distance from well bottom
(0–100 mm)
"Performing a task using Bravo Diagnostics" on
Description
The option to empty all liquid from the tips instead
of using the dispense volume specification.
The volume of liquid to be dispensed from each
pipette tip.
Specifies the volume of air to dispense after the
main volume has been dispensed while the tips
are still in the wells.
Typically, the blowout volume is the same as the
pre-aspirate volume.
Note: Blowout only occurs in the last quadrant
dispensed for a given Dispense task.
The liquid class associated with this liquid.
The distance between the end of the pipette tips
and the well bottoms during the Dispense task.
If you specify dynamic tip retraction, this is the
starting distance.
IMPORTANT
The labware definition must be
accurate and the teachpoint must be precise in
order for the system to position the tips at the
correct distance from the well bottom.
Bravo Platform User Guide