Connecting To House Vacuum With The Pinch-Valve Module - Agilent Technologies Bravo Platform User Manual

Hide thumbs Also See for Bravo Platform:
Table of Contents

Advertisement

A Accessories and platepads
Installing and setting up a Vacuum Filtration Station

Connecting to house vacuum with the pinch-valve module

120
3
Connect the Vacuum Filtration Station as required for the type of vacuum source:
House vacuum with pinch-valve module.
Agilent VARIO Pump.
Perform the following tasks to connect the pinch-valve module to the house vacuum
source.
Step
For this task...
1
Connect the manual vacuum
pressure control-valve-and-gauge
assembly.
2
Connect the pinch-valve tubing
and cables.
3
Set the pressure using the
manual vacuum control-valve
and gauge.
Connecting the control-valve-and-gauge assembly
Refer to the following figure to connect the manual pressure control-valve-and-gauge
assembly.
To connect the pressure control-valve-and-gauge assembly:
1
Ensure that the base (manifold) of the Vacuum Filtration Station is installed on the
Bravo platepad.
2
Connect the coupling insert on the end of the four-way connector (1) to the quick-
disconnect fitting (2) on the base. Push the coupling insert until it clicks.
3
Connect the coupling insert on the end of the manual vacuum control valve tubing
(3) to the quick-disconnect fitting on the short arm of the four-way connector.
4
Connect the coupling insert on the end of the vacuum pressure gauge tubing (4) to
the quick-disconnect fitting on the long arm of the four-way connector.
See...
"Connecting the control-valve-and-
gauge assembly" on page 120
"Connecting the pinch valve tubing" on
page 121
"Setting the vacuum pressure" on
page 123
Bravo Platform User Guide