E Diagnostics quick reference
Mix task parameters
256
Parameter
Aspirate distance
(0–100 mm)
Dispense at different distance
Dispense distance
(0–100 mm)
Dynamic tip extension
(0–20 mm/µL)
Pipette technique
Perform tip touch
Description
The distance between the end of the pipette tips
and the well botttoms during the aspirate action.
IMPORTANT
The labware definition must be
accurate and the teachpoint must be precise in
order for the system to position the tips at the
correct distance from the well bottom.
The option to dispense at a pipette tip height that
is different than the aspirate distance.
Select the check box to enter a value for the
dispense distance.
The distance between the end of the pipette tips
and the well bottoms during the dispense action.
The rate at which the pipette head moves during
the Aspirate task. The software calculates the
distance over which the tips will move without
crashing.
Use dynamic tip extension to prevent spills as the
pipette tips displace the liquid.
To move the tips:
At the same rate as the volume change.
Calculate dynamic tip extension (DTE) as
follows:
DTE = (well depth)/(well vol) = 1/A, where A is
the cross-sectional area of a well with straight
walls
Faster than the volume change.
DTE > 1/A
Slower than the volume change.
DTE < 1/A
The starting and ending positions can be
calculated as follows:
(V
* DTE) + Distance
aspirated
Note: Instead of a negative aspirated volume, the
software automatically moves downward toward
the well bottom with each aspirate action.
The pipette location offset you want to use for the
Dispense task.
The list of pipette techniques are defined in the
Pipette Technique Editor.
The option to touch the pipette tip on one or more
sides of the well.
Bravo Platform User Guide
well bottom