4 Maintenance and troubleshooting
Changing the Bravo head
C AU T I O N
Adjusting VWorks settings
IMPORTANT
70
Dropping the head or bumping the barrels or probes will damage the head. If the
Bravo head is not properly secured in place, it could drop unexpectedly. Ensure
that the head is securely locked into the head mount.
After installing a different Bravo head, you must ensure that the Bravo device in the
VWorks software is linked to the correct profile.
To adjust the VWorks settings for a different Bravo head:
1
In the VWorks window, ensure the correct device file is open for this Bravo
device, and then ensure that the profile selected under Bravo Properties is
correct.
If this is the first time to use this Bravo head with this device, you must create a
profile for the new Bravo head.
2
On the side of the Bravo Platform, press the power switch to the on (I) position.
3
To initialize the Bravo Platform, click Initialize selected devices in the device file.
Bravo Platform User Guide