Testing The Vacuum Filtration Station - Agilent Technologies Bravo Platform User Manual

Hide thumbs Also See for Bravo Platform:
Table of Contents

Advertisement

A Accessories and platepads
Installing and setting up a Vacuum Filtration Station

Testing the Vacuum Filtration Station

130
Before you start, ensure that you initialize the correct profile in Bravo Diagnostics.
Testing the Pump Standard option
If you selected the Standard connection option when configuring the Vacuum Filtration
Station, use the following procedure to test the vacuum supply to the station.
To test the Standard connection for a Vacuum Filtration Station:
1
Verify that the pre-assembled Vacuum Filtration Station is in place on the target
deck location and connected via the pinch-valve module.
2
In Diagnostics, click the Configuration tab.
3
In the Accessory Configuration area, highlight Vacuum Filtration Station, and then
click Diagnose accessory.
4
In the Accessory Diagnostics dialog box, start the vacuum as follows:
a
Set the Vent valve to Closed.
b
Set the Vacuum valve to Open.
c
Click Execute.
Figure
Accessory Diagnostics dialog box for the Standard connection
5
To stop the vacuum:
Bravo Platform User Guide