Testing the Orbital Shaking Station
Bravo Platform User Guide
Note: Occasionally, a port number is reassigned for a USB-to-serial
connection if you disconnect and reconnect the accessory. If this happens,
you must verify that the correct Serial Port is set in each profile.
d
Select the Module number that indicates the position of the Orbital Shaking
Station you are configuring.
For example, select 1 if the station is connected directly to the controlling
computer. Select 2 for a station that is connected to the first station, and so on.
5
Click Finish.
6
When the configuration message appears, click Yes to initialize the accessory and
move the teachpoint to a safe height. You must set the teachpoint after completing
this procedure.
7
In the Profiles tab, click Update this profile.
Test the station operation using the following procedure, and then verify and adjust the
teachpoint for the accessory location. See
plate" on page
113.
Ensure that the profile is initialized in Bravo Diagnostics before you perform the
following procedure.
To test the Orbital Shaking Station:
1
In the Accessory Configuration area, click the Orbital Shaking Station icon, and then
click Diagnose accessory. The Bravo Accessory Diagnostics dialog box opens.
A Accessories and platepads
Installing and setting up an Orbital Shaking Station
"Adjusting the teachpoint using a teach
111