Bravo Platform User Guide
Select labware
To perform a task:
1
Select a location. You can do one of the following:
Click the location in the graphical display.
Select from the Location list.
2
Select the labware from the Labware at selected location list.
3
If the task involves two locations, repeat
For example, if you select the Assemble Vacuum task for a Vacuum Filtration
Station, you could have a collection plate at Location 1 and a filter plate at
Location 2.
4
Series III disposable-tip heads and Bravo 96AM Head only. To use one row or
column of channels instead of all channels in the head:
a
Click Set head mode.
b
In the Head Mode Selector dialog box, select the barrels, and then click OK.
Note: If you select a single row on a disposable-tip head, make sure you retract
the head stripper pins.
5
If you are using the head in serial mode or the head has fewer tips than the plate
has wells, select the quadrant or quadrants in the Well Selection and Head Mode
area.
Click a representative well in the plate graphic to select the corresponding
quadrant of wells. The selection message appears below the plate graphic.
6
In the Command Parameters area:
a
Select the process from the Command to execute list.
b
Set the parameter values for the command.
5 Using diagnostics software
Performing a task using Bravo Diagnostics
step 1
and
step
2.
101