Agilent Technologies Bravo Platform User Manual page 50

Hide thumbs Also See for Bravo Platform:
Table of Contents

Advertisement

Verifying teachpoints for disposable-tip heads
Bravo Platform User Guide
The VWorks software calculates the teachpoints for the other eight locations,
keeping the z-axis coordinate the same and changing the x and y coordinates
according to their spacing.
4
Readjust the z-axis height for any positions that are taller than a standard platepad,
such as the Orbital Shaking Station.
5
To save the teachpoints, click the Profiles tab and click Update this profile.
Verify the following:
The labware definition for the tip box contains the correct column-wise and row-
wise teachpoint-to-well values.
The teachpoint is accurate for proper tips-on and tips-off operations.
To verify the tip box labware definition:
1
Open the Labware Editor.
2
In the Pipette/Well Definition tab, verify the following values for the tip box labware
definition:
Large transfer (LT) tips
Row-wise teachpoint to well
(mm) = 0
Column-wise teachpoint to well
(mm) = 0
2 Setting up the Bravo Platform
Small transfer (ST) tips
Row-wise teachpoint to well (mm) = 2.25
Column-wise teachpoint to well (mm) = 2.25
Setting teachpoints
39