C AU T I O N
Bravo Platform User Guide
b
Place your right hand under the flat area just to the left of the barrel or
syringe array to support the weight of the head.
5
While supporting the head with your hands, slide the head to the left and out of
the Bravo head mount.
Support the head securely without touching the barrels, probes, or pins. Dropping
the head or bumping the barrels, probes, or pins will damage the head.
6
Carefully place the head into the storage stand to protect the barrels, probes, as
follows:
Disposable-tip heads. Rest the bottom of the stand on a clean, dry surface.
Slide the head into the stand, with the barrels facing down, as the following
figure shows.
Figure
Series III disposable-tip head and storage stand
AssayMAP Bravo 96AM Head. Ensure that the top of the head is resting on a
clean, stable surface so that the probes are facing up.
Carefully place the stand (1) onto the head, guiding the side cutouts onto the
two side tabs (2) on the head. Use care to avoid touching the probes.
4 Maintenance and troubleshooting
Changing the Bravo head
65