4 Maintenance and troubleshooting
Troubleshooting hardware problems
Problem
The liquid in the wash trays or
reservoirs is overflowing causing
flooding.
Related information
84
Potential cause
The tubing is kinked or the Weigh
Station might require recalibration.
For information about...
Laboratory requirements and safety
guidelines
Stopping in an emergency
Reporting a problem
Solution
Inspect the tubing. If necessary,
recalibrate the Weigh Station.
See...
G5562A, G5563A Bravo Platform Safety
and Installation Guide
"Emergency stops" on page 56
"Reporting problems" on page 87
Bravo Platform User Guide