Agilent Technologies Bravo Platform User Manual page 29

Hide thumbs Also See for Bravo Platform:
Table of Contents

Advertisement

2 Setting up the Bravo Platform
Starting up and shutting down
Shutting down the Bravo Platform
C AU T I O N
Related information
18
Shut down the Bravo Platform before you:
Clean the Bravo Platform
Change the liquid-handling head
Install accessories
Move the Bravo Platform
Always turn off the Bravo Platform before installing or uninstalling a liquid-handling
head. Failure to do so can damage the head electronics.
To shut down the Bravo Platform:
1
If removing the liquid-handling head, ensure that the disposable tips are removed.
2
Optionally, home the liquid-handling head.
3
Shut down the computer.
4
Turn off any accessories, for example, Pump Module.
5
If using an Auto Filling Reservoir, disconnect the bottles to prevent siphoning.
6
On the side of the Bravo Platform, press the power switch to the off (o) position.
7
If applicable, uninstall the liquid-handling head.
For information about...
Initializing the Bravo device
Turning on or off Bravo accessories
Setting up the Bravo Platform
See...
"Initializing the device" on page 31
"Bravo accessories overview" on page 104
"Workflow for setting up the Bravo
Platform" on page 16