(Bravo Srt Only) Setting Up An Lt Tip Box Location - Agilent Technologies Bravo Platform User Manual

Hide thumbs Also See for Bravo Platform:
Table of Contents

Advertisement

A Accessories and platepads

(Bravo SRT only) Setting up an LT tip box location

(Bravo SRT only) Setting up an LT tip box location
About this topic
Before you start
C AU T I O N
172
Read this topic if you have a Bravo SRT Platform and you are using the 250-µL tips. The
height limitation on the Bravo SRT requires a special platepad and a deck location
configured for the 250-µL tip box operations.
The following figure shows the SRT platepad options for tip operations.
Figure
A
SRT platepads: (
) ST tip loading station, and (
A
To prevent potential hardware crashes when using 250-µL tips, make sure you
understand the following limitations of the z-axis clearance on the Bravo SRT.
Be aware that the Bravo SRT has insufficient clearance to move labware if the 250-
µL tips are installed on the liquid-handling head. The Bravo SRT can move labware
only if 250-µL tips are not on the liquid-handling head.
Limit tips on and tips off tasks to deck locations where an SRT 250-µL tip box
platepad is physically installed and configured in the Bravo Diagnostics software.
This special platepad provides additional z-axis clearance, and is designed for a tip
box that holds 96 250-µL tips.
Use labware classes for each deck location so that the 250-µL tip box operations
are not attempted at deck locations with insufficient z-axis clearance.
Use VWorks Task Groups to ensure that tips on and tips off tasks are performed
before plate movement tasks in the protocol.
Ensure that you have the following:
M5 hex wrench for removing a standard SRT platepad
SRT platepad for 250-µL tip box
Star-head screw and wrench for installing the SRT 250-µL tip box platepad
B
) SRT platepad for 250 µL LT tip boxes
B
Bravo Platform User Guide