Dispense To Waste Task Parameters - Agilent Technologies Bravo Platform User Manual

Hide thumbs Also See for Bravo Platform:
Table of Contents

Advertisement

E Diagnostics quick reference

Dispense to Waste task parameters

Related information

Dispense to Waste task parameters

About this topic
Parameter descriptions
252
For information about...
Opening Diagnostics
Editing the labware classes, liquid
library, or pipetting techniques
Performing a diagnostic task
Reporting a problem
This topic describes the task parameters for the Dispense to Waste task, which is
available only for a location with a Tip Wash Station (also known as a MicroWash
Reservoir).
The Dispense to Waste task moves the pipettes by a horizontal offset and then
dispenses used fluid in between the chimneys in the Tip Wash Station. The software
calculates the horizontal offset automatically based on the labware definition for the
Tip Wash Station.
To run the Dispense to Waste task from Diagnostics, see
Bravo Diagnostics" on page
Automation Control User
The Dispense to Waste task has the following parameters.
Parameter
Empty tips
Volume (µL)
Blowout volume (µL)
Liquid class
See...
"Opening Bravo Diagnostics" on page 24
VWorks Automation Control Setup Guide
"Performing a task using Bravo Diagnostics"
on page 100
"Reporting problems" on page 87
100. To use the task in a protocol, see the
Guide.
Description
The option to empty all the liquid from the head
into the Tip Wash Station outside the chimneys.
The volume of liquid to be dispensed from each
pipette.
Specifies the volume of air to dispense after the
main volume has been dispensed while the tips
are still in the Tip Wash Station.
The pipetting speed and accuracy.
IMPORTANT
always select a liquid class for liquid-handling
tasks.
"Performing a task using
VWorks
To ensure consistent pipetting,
Bravo Platform User Guide