Parameter
Pump fill speed (%)
Pump empty speed (%)
Dispense to waste during wash
Dispense to waste at height
(mm)
For information about...
Opening Diagnostics
Editing the labware classes, liquid
library, or pipetting techniques
Performing a diagnostic task
Reporting a problem
E Diagnostics quick reference
Wash Tips task parameters
Description
The speed, in percent of maximum speed, of
liquid flow into the reservoir.
For the MicroWash Reservoir, this value should be
high enough for the washing liquid to just bubble
over the tops of the chimneys.
The speed, in percent of maximum speed, of
liquid flow out of the the reservoir.
For the MicroWash Reservoir, this value should be
slightly higher than that of the inflow pump to
prevent an overflow.
The option to move the tips by a specified offset
(defined in the Labware Editor) and dispense used
fluid outside of the reservoir chimney.
This option applies only to reservoirs that have
chimneys.
The height at which the dispense action occurs.
For example, during the dispense action, the tips
move up to clear the chimneys, move the offset
distance, and then lower to the distance you
specified. If you want the lower the tips by 10 mm,
specify –10 mm.
See...
"Opening Bravo Diagnostics" on page 24
VWorks Automation Control Setup Guide
"Performing a task using Bravo Diagnostics"
on page 100
"Reporting problems" on page 87