Configuring a platepad for delidding

About this topic
Before you start
Configuring a platepad for delidding
Related information
Bravo Platform User Guide
This topic explains how to use Bravo Diagnostics to configure a platepad to accept a lid
in a delidding process.
Make sure you install a platepad at the desired deck location.
Do not install another accessory at this deck location.
To configure a platepad for delidding:
1
In Diagnostics, click the Profiles tab, and verify that the correct profile is initialized.
2
Click the Configuration tab.
3
In the Location is configured as list, select Accessory. The Accessories Wizard
appears.
4
Follow the instructions in the Accessories Wizard to:
Choose the Location on which you want to place the microplate lid.
Note: Your selection in the Accessories Wizard overrides the selection in the
Configuration tab.
Select the Lid hotel accessory to indicate that you want to use the Delidding
and Relidding tasks.
Note: Because you can place only one lid on the designated platepad, Number of
Slots is set to 1.
5
When the accessory configuration message appears, do one of the following:
If you have not yet edited the teachpoint, click Yes. After the software initializes
the accessory, edit the teachpoint.
If you have already verified the teachpoint, click No.
6
In the Profiles tab, click Update this profile.
For information about...
Opening Diagnostics
Editing teachpoints
Using an accessory in a protocol
A Accessories and platepads
Configuring a platepad for delidding
See...
"Opening Bravo Diagnostics" on page 24
"Setting teachpoints" on page 34
VWorks Automation Control User Guide