Parameter
Tip touch horizontal offset
(–5 to 5 mm)
For information about...
Opening Diagnostics
Editing the labware classes, liquid
library, or pipetting techniques
Performing a diagnostic task
Reporting a problem
E Diagnostics quick reference
AM Aspirate task parameters
Description
The horizontal distance the tips move. The value
is based on the well diameter specified by the
labware definition.
For example, if you set a value of:
0, the tips move a horizontal distance equal to
the well radius
> 0, the tips attempt to move past the well
radius, which results in a more forceful tip
touch
< 0, the tips move a distance less than the
radius of the well, resulting in a lighter tip
touch
See...
"Opening Bravo Diagnostics" on page 24
VWorks Automation Control Setup Guide
"Performing a task using Bravo Diagnostics"
on page 100
"Reporting problems" on page 87