E Diagnostics quick reference

Wash Tips task parameters

Related information

Wash Tips task parameters

About this topic
Parameter descriptions
264
Parameter
Target pressure
ME4C VARIO Vacuum Pump only
Vent delay
ME4C VARIO Vacuum Pump only
For information about...
Opening Diagnostics
Editing the labware classes, liquid
library, or pipetting techniques
Setting up the Vacuum Filtration
Station
Performing a diagnostic task
Reporting a problem
The Wash Tips task washes pipette tips using a number of aspirate and dispense
actions. This topic describes the task parameters.
To run the Wash Tips task, see
page
100.
The Wash Tips task has the following parameters.
Description
The difference between the pressure of the
outside atmosphere above the filter and the
pressure in the Vacuum Filtration Station
manifold, including the enclosure beneath the
filter.
For example, if you set the Target pressure to
600 mbar and the ambient pressure displayed
on the Vario pump is 1000 mbar, the vacuum
will remain on until the reading on the Vario
pump reaches 400 mbar.
The length of time, in seconds, to wait for the
air pressure under the filter to equalize with the
ambient air pressure.
See...
"Opening Bravo Diagnostics" on page 24
VWorks Automation Control Setup Guide
"Installing and setting up a Vacuum
Filtration Station" on page 114
"Performing a task using Bravo Diagnostics"
on page 100
"Reporting problems" on page 87
"Performing a task using Bravo Diagnostics" on
Bravo Platform User Guide