D AssayMAP Bravo Platform
Requirements for partial-head liquid handling
Requirements for mounting a partial head of 250-µL pipette tips
C AU T I O N
C AU T I O N
Related information
222
Although the VWorks software considers the geometry of the labware at the target deck
location during the task, the software is unaware of the labware height on the adjacent
deck location where the head will be offset.
A collision will result if the Bravo 96AM Head is offset from the tip box location when
pressing on pipette tips. When picking up pipette tips from a tip box, ensure that the
Bravo 96AM Head is centered over the tip box.
To ensure consistent seating of pipette tips on the probes, use the 96AM Cartridge &
Tip Seating Station to seat the pipette tips on the probes.
The Bravo 96AM Head must be centered over the tip box when picking up pipette tips to
be transfered to the 96AM Cartridge & Tip Seating Station.
Figure
96AM Cartridge & Tip Seating Station
For more information about...
Bravo 96AM Head
Using and storing the Bravo 96AM
Head
See...
"Bravo 96AM Head overview" on page 200
"Installing and using the Bravo 96AM Head"
on page 208