Agilent Technologies 8890 Maintaining Manual

Agilent Technologies 8890 Maintaining Manual

Gas chromatograph
Hide thumbs Also See for 8890:
Table of Contents

Advertisement

Quick Links

Agilent 8890 Gas Chromatograph
Maintaining Your Gas Chromatograph (GC)

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 8890

  • Page 1 Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph (GC)
  • Page 2 Notices Warranty Safety Notices © Agilent Technologies, Inc. 2019 No part of this manual may be reproduced in The material contained in this document is any form or by any means (including provided “as is,” and is subject to being...
  • Page 3: Table Of Contents

    Contents 1 About Maintaining the GC Overview of Maintenance Tools and Materials Required for Maintenance Preparing for Maintenance Viewing Status Data Finding a Replacement Part Number 2 Removing Covers To Remove the Detector Top Cover To Remove the Pneumatics Cover To Remove the Electronics Cover 3 Maintaining the GC Consumables and Parts for General GC Maintenance...
  • Page 4 To Change the Liner and O-Ring on the Split/Splitless Inlet To Replace the Gold Seal on the Split/Splitless Inlet To Replace the Filter in the Split Vent Trap for the Split/Splitless Inlet To Clean the Split/Splitless Inlet To Bakeout Contaminants from the Split/Splitless Inlet 6 Maintaining the Purged Packed Inlet Consumables and Parts for the Purged Packed Inlet Exploded Parts View of the Purged Packed Inlet...
  • Page 5 To Replace the Fused Silica Needle in a Syringe for the COC Inlet To Bakeout Contaminants from the COC Inlet 8 Maintaining the MMI Consumables and Parts for the Multimode Inlet Exploded Parts View of the MMI To Install the Column Nut Adapter To Install a Capillary Column with the Multimode Inlet To Change the Septum on the MMI To Clean the Septum Seat in the Insert Assembly of the MMI...
  • Page 6 To Clean the VI To Install the VI Interface To Replace the Filter in the Split Vent Trap for the VI To Bakeout Contaminants from the VI Inlet 11 Maintaining the FID Consumables and Parts for the FID Exploded Parts Views of the Flame Ionization Detector To Install a Capillary Column in the FID To Replace an FID Jet To Perform Maintenance on the FID Collector Assembly...
  • Page 7 To Replace the ECD Fused Silica Indented Mixing Liner and Install the Makeup Gas Adapter To Install a Capillary Column in the ECD To Bakeout the ECD 15 Maintaining the FPD+ Consumables and Parts for the FPD+ Exploded Parts View of the Flame Photometric Detector Plus To Install a Capillary Column to the FPD+ To Change the FPD+ Wavelength Filter To Remove the FPD+ Cover...
  • Page 8 Maintaining Your GC...
  • Page 9: About Maintaining The Gc

    About Maintaining the GC Overview of Maintenance 10 Tools and Materials Required for Maintenance 11 Preparing for Maintenance 13 Viewing Status Data 14 Finding a Replacement Part Number 16 This section provides an overview of the maintenance procedures included in this document. It also lists the tools needed for routine maintenance and the information needed to safely perform the various maintenance task.
  • Page 10: Overview Of Maintenance

    Overview of Maintenance Overview of Maintenance This manual details the routine tasks needed to maintain the Agilent 8890 Gas Chromatograph (GC). The procedures assume a basic knowledge of tool use and of GC operation. Readers are, for example, expected to know how to: •...
  • Page 11: Tools And Materials Required For Maintenance

    About Maintaining the GC Tools and Materials Required for Maintenance Tools and Materials Required for Maintenance Table 1 lists the tools needed for most GC maintenance procedures. The specific tools required to perform a maintenance procedure are listed in step 1 of each procedure. Table 1 Tools and materials for GC maintenance Common tools...
  • Page 12 About Maintaining the GC Tools and Materials Required for Maintenance Table 1 Tools and materials for GC maintenance Common tools Part number Gloves, clean, lint-free, nylon (for handling contamination-sensitive parts) large: 8650-0030 small: 8650-0029 Steel wool, 0- or 00-grade (for cleaning an inlet’s septum seating surfaces) * Included with the GC ship kits Maintaining Your GC...
  • Page 13: Preparing For Maintenance

    About Maintaining the GC Preparing for Maintenance Preparing for Maintenance Before routine maintenance procedures, the GC must be made ready. This process can include: • Setting low temperatures to avoid burns and other injuries • Setting reduced flows to avoid safety hazards and to prevent damage to the instrument •...
  • Page 14: Viewing Status Data

    About Maintaining the GC Viewing Status Data Viewing Status Data The GC can display signal and other data that can be very useful during maintenance. For example: • Instrument status: Readiness state, running state. • Instrument status details: Not ready modules, error modules, warnings. •...
  • Page 15 About Maintaining the GC Viewing Status Data Select the desired item from the drop-down list, then touch Add. Maintaining Your GC...
  • Page 16: Finding A Replacement Part Number

    About Maintaining the GC Finding a Replacement Part Number Finding a Replacement Part Number Agilent now provides the Parts Finder tool to help you locate replacement and consumable part numbers. If using an Agilent data system, Parts Finder will be installed. If you want to install the tool on another computer, Parts Finder is included on the Agilent GC and GC/MS User Manuals &...
  • Page 17: Removing Covers

    Removing Covers To Remove the Detector Top Cover 18 To Remove the Pneumatics Cover 19 To Remove the Electronics Cover 20 This section describes how to remove covers as needed for routine maintenance. Only the covers listed in this chapter should be removed. Removing other GC covers can compromise the safety features of the GC, leading to personal injury or damage to the instrument.
  • Page 18: To Remove The Detector Top Cover

    Do not force the cover, either when installing it or closing it. This could break the plastic C AU T I O N parts. Pneumatics cover Detector cover Pneumatics cover release button (on both sides) Injection port cover Left side cover Power switch Figure 2. 8890 GC covers Maintaining Your GC...
  • Page 19: To Remove The Pneumatics Cover

    Removing Covers To Remove the Pneumatics Cover To Remove the Pneumatics Cover The pneumatics cover protects the flow manifolds in the back top of the GC. 1 Disconnect any vent tubing connected to the split and septum purge vents. 2 Press the buttons located on each side of the pneumatics cover and lift up to remove it from the GC.
  • Page 20: To Remove The Electronics Cover

    With FPD+ installed: 1 Turn off the GC and unplug the power cord. 2 Raise or remove the detector top cover. Loosen screw Electronics cover Figure 3. 8890 GC electronics cover Maintaining Your GC...
  • Page 21 Removing Covers To Remove the Electronics Cover Loosen captive screws on right side cover Pneumatics cover Electronics cover release button (on both sides) GC front GC back Figure 4. Loosen captive screws located on the right side of the GC 3 Loosen the two captive screws on the right side cover, slide the cover backwards, then remove.
  • Page 22 Removing Covers To Remove the Electronics Cover Maintaining Your GC...
  • Page 23: Maintaining The Gc

    Maintaining the GC Consumables and Parts for General GC Maintenance Parts Identification To Change the Hydrogen Sensor Calibration Gas Cylinder Maintaining Your GC...
  • Page 24: Consumables And Parts For General Gc Maintenance

    Maintaining the GC Consumables and Parts for General GC Maintenance Consumables and Parts for General GC Maintenance Table 2 lists consumable parts for the hydrogen sensor module accessory and for general GC maintenance. Table 2 Consumables and parts for general GC maintenance Description Part number Hydrogen sensor parts...
  • Page 25 Part number Gas Clean connecting unit, 2-position), 1/8-inch CP738407 Gas Clean Smart Sensor for 8890 (replacement sensor) CP179885 Gas Clean kit (connecting unit for one filter, one carrier gas filter, 1/8-inch connections, mounting bracket, and Smart Sensor) for new GC...
  • Page 26: Parts Identification

    Maintaining the GC Parts Identification Parts Identification 1/8-inch Swagelok nut and ferrule Pressure regulator Calibration gas 1/8-inch copper tubing, cylinder cut to length needed Gas cylinder stand Figure 5. Hydrogen sensor calibration gas cylinder parts identification Maintaining Your GC...
  • Page 27: To Change The Hydrogen Sensor Calibration Gas Cylinder

    Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder To Change the Hydrogen Sensor Calibration Gas Cylinder Replace the calibration gas cylinder when you cannot adjust it to provide the correct flow rate during calibration, or if it expires. Connect only the Agilent calibration gas to the hydrogen sensor fitting.
  • Page 28 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder Maintaining Your GC...
  • Page 29: Maintaining Capillary Columns

    Maintaining Capillary Columns Consumables and Parts for Columns 30 To Install a Capillary Column Hanger 32 To Install Capillary Column Clips 33 To Condition a Capillary Column 34 To Cut a Loop from a Column 36 To Reverse a Column and Bakeout Contaminants 37 To Attach a Capillary Column Using UltiMetal Plus Flexible Metal Fittings 39 To Remove the Ferrule From a CFT Fitting 47 Maintaining Your GC...
  • Page 30: Consumables And Parts For Columns

    Maintaining Capillary Columns Consumables and Parts for Columns Consumables and Parts for Columns See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 3 Nuts, ferrules, and hardware for capillary columns Column id (mm) Description Typical use...
  • Page 31 Maintaining Capillary Columns Consumables and Parts for Columns Table 4 Capillary column hangers Description Part Number Column hanger 1460-1914 Capillary column clip kit, for 7-inch column basket G1530-61580 Maintaining Your GC...
  • Page 32: To Install A Capillary Column Hanger

    Maintaining Capillary Columns To Install a Capillary Column Hanger To Install a Capillary Column Hanger Be careful! The oven may be hot enough to cause burns. If the oven is hot, wear WAR N IN G heat-resistant gloves to protect your hands. Wear safety glasses to protect your eyes from flying particles while handling, cutting, or WAR N IN G installing glass or fused silica capillary columns.
  • Page 33: To Install Capillary Column Clips

    Maintaining Capillary Columns To Install Capillary Column Clips To Install Capillary Column Clips Be careful! The oven may be hot enough to cause burns. If the oven is hot, wear WAR N IN G heat-resistant gloves to protect your hands. Wear safety glasses to protect your eyes from flying particles while handling, cutting, or WAR N IN G installing glass or fused silica capillary columns.
  • Page 34: To Condition A Capillary Column

    Maintaining Capillary Columns To Condition a Capillary Column To Condition a Capillary Column 1 Gather the following: • One 7/16-inch, and 1/4-inch wrenches • No-hole ferrule. See “Consumables and Parts for Columns” on page 30. • Column nut Do not use hydrogen as the carrier for conditioning! It could vent into the oven and present an WAR N IN G explosion hazard.
  • Page 35 Maintaining Capillary Columns To Condition a Capillary Column 6 Turn off all detectors. 7 Set a minimum velocity of 30 cm/s, or as recommended by the column manufacturer. Let gas flow through the column at room temperature for 15 to 30 minutes to remove air. 8 Program the oven from room temperature to the maximum temperature limit for the column.
  • Page 36: To Cut A Loop From A Column

    Maintaining Capillary Columns To Cut a Loop from a Column To Cut a Loop from a Column 1 Gather the following: • New ferrule(s) for the column inlet connection • Column cutter 2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance > Maintenance Mode >...
  • Page 37: To Reverse A Column And Bakeout Contaminants

    Maintaining Capillary Columns To Reverse a Column and Bakeout Contaminants To Reverse a Column and Bakeout Contaminants 1 Gather the following: • 1/4-inch wrench • Column cutter 2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance > Maintenance Mode >...
  • Page 38 Maintaining Capillary Columns To Reverse a Column and Bakeout Contaminants 11 Set the inlet temperature to 300 °C or 25 °C above the normal operating temperature. 12 Set the column oven 25 °C above the GC method final oven temperature to bake contaminants out of the inlet, mostly through the split vent.
  • Page 39: To Attach A Capillary Column Using Ultimetal Plus Flexible Metal Fittings

    Maintaining Capillary Columns To Attach a Capillary Column Using UltiMetal Plus Flexible Metal Fittings To Attach a Capillary Column Using UltiMetal Plus Flexible Metal Fittings The available UltiMetal Plus Flexible Metal ferrule packages are listed in Table 5 Figure Table 5 Available UltiMetal Plus Flexible Metal ferrule packages Item Part number...
  • Page 40 Maintaining Capillary Columns Preparing the column installation to the CFT fitting Figure 7. Thread the column through the internal nut, ferrule, and swaging tool 4 Insert the column into the swaging tool until it bottoms. 5 Thread the internal nut into the swaging tool until finger tight. See Figure Figure 8.
  • Page 41 Maintaining Capillary Columns Preparing the column installation to the CFT fitting Table 6 Degrees to tighten the internal nut for proper swaging Ferrule part number Degrees to tighten internal nut G3188-27501 50 – 100 degrees G3188-27502 30 – 70 degrees G3188-27503 20 –...
  • Page 42 Maintaining Capillary Columns Preparing the column installation to the inlet 8 Using the 1/4-inch open-ended wrench, remove the internal nut from the swaging tool and column. Figure 11. Properly-trimmed column and ferrule Preparing the column installation to the inlet This procedure is used to attach a capillary column to a splitter or switch or an Ultimate Union. 1 Gather the following: •...
  • Page 43 Maintaining Capillary Columns Preparing the column installation to the inlet Figure 13. Thread the column through the column nut, ferrule, and swaging tool 4 Loosely thread the column nut into the swaging tool until you start to feel resistance. The column should still slide freely.
  • Page 44 Maintaining Capillary Columns Preparing the column installation to the inlet Figure 15. Align column (split/splitless shown) 6 Begin to swage the ferrule to the column using a 1/4-inch open-ended wrench. The design of UltiMetal Plus Flexible Metal ferrules reduces the probability of column breakage and minimizes the damage to the thread of the corresponding fittings.
  • Page 45 Maintaining Capillary Columns Preparing the column installation to the inlet Figure 16. Tighten the column nut 7 Using the 1/4-inch open-ended wrench, turn the column nut an additional 15 to 20 degrees clockwise to assure the ferrule is properly swaged onto the column. See Figure Figure 17.
  • Page 46 Maintaining Capillary Columns Swaging Quality Inspection Swaging Quality Inspection Figure 18 shows both an incorrect and correct swaging symmetry. Incorrect Correct Figure 18. Incorrect and correct swaging symmetry If your column and ferrule appear like the incorrect example in Figure 18 above, it is possible that your swaging wrench is defective or worn out.
  • Page 47: To Remove The Ferrule From A Cft Fitting

    Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting To Remove the Ferrule From a CFT Fitting Loosen and remove the internal nut. If the ferrule does not release from the fitting, do not try to pull the column free. Insert a pointed object (such as a thumb pin or a paper clip) into the ferrule release hole, as shown Figure...
  • Page 48 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting Maintaining Your GC...
  • Page 49: Maintaining The Split/Splitless Inlet

    Maintaining the Split/Splitless Inlet Consumables and Parts for the Split/Splitless Inlet 50 Exploded Parts View of the Split/Splitless Inlet 53 To Install a Capillary Column with the Split/Splitless Inlet 54 To Change the Septum on the Split/Splitless Inlet 58 To Clean the Septum Seat in the Insert Assembly of the Split/Splitless Inlet 60 To Change the Liner and O-Ring on the Split/Splitless Inlet 62 To Replace the Gold Seal on the Split/Splitless Inlet 65 To Replace the Filter in the Split Vent Trap for the Split/Splitless Inlet 67...
  • Page 50: Consumables And Parts For The Split/Splitless Inlet

    Maintaining the Split/Splitless Inlet Consumables and Parts for the Split/Splitless Inlet Consumables and Parts for the Split/Splitless Inlet See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 8 Split, splitless, direct, and direct connect inlet liners Mode...
  • Page 51 Maintaining the Split/Splitless Inlet Consumables and Parts for the Split/Splitless Inlet Table 9 Nuts, ferrules, and hardware for capillary columns (continued) Column id (mm) Description Typical use Part number/quantity .250 Ferrule, Vespel/graphite, 0.4-mm id 0.1-mm, 0.2-mm, and 0.25-mm capillary 5181-3323 (10/pk) columns Ferrule, graphite, 0.5-mm id 0.1-mm, 0.2-mm, 0.25-mm, and 0.32-mm...
  • Page 52 Maintaining the Split/Splitless Inlet Consumables and Parts for the Split/Splitless Inlet Table 10 Other consumables and parts for the split/splitless inlet (continued) Description/quantity Part number Column nut, blanking plug 5020-8294 Capillary inlet preventative maintenance kit, split 5188-6496 Capillary inlet preventative maintenance kit, splitless 5188-6497 Maintaining Your GC...
  • Page 53: Exploded Parts View Of The Split/Splitless Inlet

    Maintaining the Split/Splitless Inlet Exploded Parts View of the Split/Splitless Inlet Exploded Parts View of the Split/Splitless Inlet Septum retainer nut Merlin cap Septum Merlin Microseal Insert assembly O-ring Liner Split vent line Split/splitless inlet body Retaining nut Inlet gold seal Washer Reducing nut Insulation...
  • Page 54: To Install A Capillary Column With The Split/Splitless Inlet

    Maintaining the Split/Splitless Inlet To Install a Capillary Column with the Split/Splitless Inlet To Install a Capillary Column with the Split/Splitless Inlet Do not use hydrogen as the carrier for conditioning! It could vent into the oven and present an WAR N IN G explosion hazard.
  • Page 55 Maintaining the Split/Splitless Inlet To Install a Capillary Column with the Split/Splitless Inlet Ferrule Column nut Septum 6 Score the column using a glass scribing tool. The score must be square to ensure a clean break. 7 Break off the column end by supporting it against the column cutter opposite the scribe. Inspect the end with a magnifying loupe to make certain there are no burrs or jagged edges.
  • Page 56 Maintaining the Split/Splitless Inlet To Install a Capillary Column with the Split/Splitless Inlet 4%6 mm 10 Thread the column nut into the inlet but do not tighten. 11 Adjust the column position so that the septum contacts the bottom of the column nut. Finger-tighten the column nut until it begins to grip the column.
  • Page 57 Maintaining the Split/Splitless Inlet To Install a Capillary Column with the Split/Splitless Inlet 16 Install the column into the detector. • To Install a Capillary Column in the FID • To Install a Capillary Column in the NPD • To Install a Capillary Column in the TCD •...
  • Page 58: To Change The Septum On The Split/Splitless Inlet

    Maintaining the Split/Splitless Inlet To Change the Septum on the Split/Splitless Inlet To Change the Septum on the Split/Splitless Inlet 1 Gather the following: • Replacement septum, see “Consumables and Parts for the Split/Splitless Inlet” page 50. • Wrench, hex for changing septum •...
  • Page 59 Maintaining the Split/Splitless Inlet To Change the Septum on the Split/Splitless Inlet 1 mm 7 Select Maintenance > Inlets > Septum injections, then touch Reset Counter. 8 Perform an inlet Leak & Restriction test and reset the maintenance counters. 9 Select Finished, then select OK to exit the GC maintenance wizard. 10 Restore the analytical method.
  • Page 60: To Clean The Septum Seat In The Insert Assembly Of The Split/Splitless Inlet

    Maintaining the Split/Splitless Inlet To Clean the Septum Seat in the Insert Assembly of the Split/Splitless Inlet To Clean the Septum Seat in the Insert Assembly of the Split/Splitless Inlet 1 Gather the following: • Replacement septum, see “Consumables and Parts for the Split/Splitless Inlet” page 50.
  • Page 61 Maintaining the Split/Splitless Inlet To Clean the Septum Seat in the Insert Assembly of the Split/Splitless Inlet 6 Scrub the residue from the retainer nut and septum holder with a small piece of rolled-up steel wool and tweezers. Do not do this over the inlet. 7 Use compressed air or nitrogen to blow away the pieces of steel wool and septum.
  • Page 62: To Change The Liner And O-Ring On The Split/Splitless Inlet

    Maintaining the Split/Splitless Inlet To Change the Liner and O-Ring on the Split/Splitless Inlet To Change the Liner and O-Ring on the Split/Splitless Inlet 1 Gather the following: • Replacement O-ring, see “Consumables and Parts for the Split/Splitless Inlet” page 50. •...
  • Page 63 Maintaining the Split/Splitless Inlet To Change the Liner and O-Ring on the Split/Splitless Inlet 5 Grasp the liner with tweezers and pull it out. 6 Inspect the surface of the gold seal for graphite or rubber septum contamination. If required, replace the gold seal. See“To Replace the Gold Seal on the Split/Splitless Inlet”...
  • Page 64 Maintaining the Split/Splitless Inlet To Change the Liner and O-Ring on the Split/Splitless Inlet 7 Clean the inlet if there is visible or suspected contamination. See “To Clean the Split/Splitless Inlet” on page 69. 8 Clean O-ring residue from sealing surface. Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
  • Page 65: To Replace The Gold Seal On The Split/Splitless Inlet

    Maintaining the Split/Splitless Inlet To Replace the Gold Seal on the Split/Splitless Inlet To Replace the Gold Seal on the Split/Splitless Inlet 1 Gather the following: • Replacement gold seal, see “Consumables and Parts for the Split/Splitless Inlet” page 50. •...
  • Page 66 Maintaining the Split/Splitless Inlet To Replace the Gold Seal on the Split/Splitless Inlet 6 Put on gloves to protect the new gold seal and washer from contamination. Put a new washer in the reducing nut and place the new gold seal on top of it (raised portion facing down).
  • Page 67: To Replace The Filter In The Split Vent Trap For The Split/Splitless Inlet

    Maintaining the Split/Splitless Inlet To Replace the Filter in the Split Vent Trap for the Split/Splitless Inlet To Replace the Filter in the Split Vent Trap for the Split/Splitless Inlet 1 Gather the following: • New split vent filter cartridge. See “Consumables and Parts for the Split/Splitless Inlet”...
  • Page 68 Maintaining the Split/Splitless Inlet To Replace the Filter in the Split Vent Trap for the Split/Splitless Inlet Filter and O-ring Knurled nut Figure 23. Slide trap assembly to expose the filter 6 Remove the old filter cartridge and two O-rings. 7 Verify the new O-rings are seated properly on the new filter cartridge.
  • Page 69: To Clean The Split/Splitless Inlet

    Maintaining the Split/Splitless Inlet To Clean the Split/Splitless Inlet To Clean the Split/Splitless Inlet 1 Gather the following: • Replacement septum, see “Consumables and Parts for the Split/Splitless Inlet” page 50. • Replacement liner • Replacement O-ring • Replacement gold seal •...
  • Page 70 Maintaining the Split/Splitless Inlet To Clean the Split/Splitless Inlet 13 Perform an inlet Leak & Restriction test and reset the maintenance counters. 14 Bakeout contaminants. See “To Bakeout Contaminants from the Split/Splitless Inlet” page 71. 15 On the GC touchscreen, select Maintenance > Inlets, then touch Reset Counter. 16 Exit maintenance mode.
  • Page 71: To Bakeout Contaminants From The Split/Splitless Inlet

    Maintaining the Split/Splitless Inlet To Bakeout Contaminants from the Split/Splitless Inlet To Bakeout Contaminants from the Split/Splitless Inlet 1 Put the inlet into split mode. 2 Set the column flow to the normal operating value, or set the capillary column gas velocity to 30 cm/s.
  • Page 72 Maintaining the Split/Splitless Inlet To Bakeout Contaminants from the Split/Splitless Inlet Maintaining Your GC...
  • Page 73: Maintaining The Purged Packed Inlet

    Maintaining the Purged Packed Inlet Consumables and Parts for the Purged Packed Inlet 74 Exploded Parts View of the Purged Packed Inlet 77 To Install a Capillary Column with the Purged Packed Inlet 78 To Change the Septum on the Purged Packed Inlet 82 To Clean the Septum Seat in the Purged Packed Inlet 84 To Install an Adapter on the Purged Packed Inlet 86 To Change the O-Ring on the Purged Packed Inlet 87...
  • Page 74: Consumables And Parts For The Purged Packed Inlet

    Maintaining the Purged Packed Inlet Consumables and Parts for the Purged Packed Inlet Consumables and Parts for the Purged Packed Inlet See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 11 Purged packed inlet parts Description Part number/quantity...
  • Page 75 Maintaining the Purged Packed Inlet Consumables and Parts for the Purged Packed Inlet Table 12 Nuts and ferrules for packed columns (continued) Description Typical use Part number/quantity 1/4-inch id Swagelok brass nut, front ferrule, back 1/4-inch column 5080-8752 (20 each/pk) ferrule 1/4-inch id Vespel/ graphite ferrule Inlet/detector liner/adapters...
  • Page 76 Maintaining the Purged Packed Inlet Consumables and Parts for the Purged Packed Inlet Table 13 Nuts, ferrules, and hardware for capillary columns (continued) Column id (mm) Description Typical use Part number/quantity Ferrule tool kit Ferrule installation 440-1000 Maintaining Your GC...
  • Page 77: Exploded Parts View Of The Purged Packed Inlet

    Maintaining the Purged Packed Inlet Exploded Parts View of the Purged Packed Inlet Exploded Parts View of the Purged Packed Inlet Septum nut Merlin cap Septum Merlin Microseal Top insert weldment O-ring Glass liner Ferrule Adapter nut Adapter Insulation Insulation cup Ferrule Column nut Figure 24.
  • Page 78: To Install A Capillary Column With The Purged Packed Inlet

    Maintaining the Purged Packed Inlet To Install a Capillary Column with the Purged Packed Inlet To Install a Capillary Column with the Purged Packed Inlet 1 Gather the following: • Column • Ferrule, see “Consumables and Parts for the Purged Packed Inlet” on page 74.
  • Page 79 Maintaining the Purged Packed Inlet To Install a Capillary Column with the Purged Packed Inlet 6 Score the column using a glass scribing tool. The score must be square to ensure a clean break. 7 Break off the column end by supporting it against the column cutter opposite the scribe. Inspect the end with a magnifying loupe to make certain there are no burrs or jagged edges.
  • Page 80 Maintaining the Purged Packed Inlet To Install a Capillary Column with the Purged Packed Inlet 11 Adjust the column position so that the septum is even with the bottom of the column nut. Finger-tighten the column nut until it begins to grip the column. 12 Tighten the column nut an additional 1/4 to 1/2 turn with a wrench so that the column cannot be pulled from the fitting with gentle pressure.
  • Page 81 Maintaining the Purged Packed Inlet To Install a Capillary Column with the Purged Packed Inlet 18 After the column is installed at both inlet and detector, establish a flow of carrier gas and purge as recommended by the column manufacturer. 19 Restore the analytical method.
  • Page 82: To Change The Septum On The Purged Packed Inlet

    Maintaining the Purged Packed Inlet To Change the Septum on the Purged Packed Inlet To Change the Septum on the Purged Packed Inlet 1 Gather the following: • Replacement septum, see “Consumables and Parts for the Purged Packed Inlet” page 74. •...
  • Page 83 Maintaining the Purged Packed Inlet To Change the Septum on the Purged Packed Inlet 6 Replace the septum retainer nut or Merlin cap and finger-tighten. Tighten the septum retainer nut until the C-ring is about 1 mm above the nut. Overtightening the septum nut can cause contamination.
  • Page 84: To Clean The Septum Seat In The Purged Packed Inlet

    Maintaining the Purged Packed Inlet To Clean the Septum Seat in the Purged Packed Inlet To Clean the Septum Seat in the Purged Packed Inlet 1 Gather the following: • Replacement septum, see “Consumables and Parts for the Purged Packed Inlet” page 74.
  • Page 85 Maintaining the Purged Packed Inlet To Clean the Septum Seat in the Purged Packed Inlet 1 mm 12 On the GC touchscreen select Maintenance > Inlets > Septum injections, then select Reset Counter. 13 Perform an inlet Leak & Restriction test and reset the maintenance counters. 14 Exit maintenance mode.
  • Page 86: To Install An Adapter On The Purged Packed Inlet

    Maintaining the Purged Packed Inlet To Install an Adapter on the Purged Packed Inlet To Install an Adapter on the Purged Packed Inlet 1 Gather the following: • Brass tubing nut, see “Consumables and Parts for the Purged Packed Inlet” page 74.
  • Page 87: To Change The O-Ring On The Purged Packed Inlet

    Maintaining the Purged Packed Inlet To Change the O-Ring on the Purged Packed Inlet To Change the O-Ring on the Purged Packed Inlet 1 Gather the following: • Replacement O-ring, see “Consumables and Parts for the Purged Packed Inlet” page 74. •...
  • Page 88: To Change The Glass Liner On The Purged Packed Inlet

    Maintaining the Purged Packed Inlet To Change the Glass Liner on the Purged Packed Inlet To Change the Glass Liner on the Purged Packed Inlet 1 Gather the following: • Replacement glass liner, see “Consumables and Parts for the Purged Packed Inlet” page 74.
  • Page 89 Maintaining the Purged Packed Inlet To Change the Glass Liner on the Purged Packed Inlet 6 Wearing gloves, grasp the flared end (top) of the replacement glass liner with tweezers and install it in the inlet. If the glass liner does not seat properly because a capillary column is installed, remove the column, install the glass liner, and replace the column.
  • Page 90: To Clean The Purged Packed Inlet

    Maintaining the Purged Packed Inlet To Clean the Purged Packed Inlet To Clean the Purged Packed Inlet 1 Gather the following: • Replacement O-ring, see “Consumables and Parts for the Purged Packed Inlet” page 74. • Replacement glass liner • Replacement septum •...
  • Page 91 Maintaining the Purged Packed Inlet To Clean the Purged Packed Inlet 15 Install the septum and septum nut. See “To Change the Septum on the Purged Packed Inlet” on page 82. 16 Attach the column. See “To Install a Capillary Column with the Purged Packed Inlet” page 78.
  • Page 92: To Bakeout Contaminants From The Purged Packed Inlet

    Maintaining the Purged Packed Inlet To Bakeout Contaminants from the Purged Packed Inlet To Bakeout Contaminants from the Purged Packed Inlet 1 Set the column flow to the normal operating value, or set the capillary column gas velocity to 30 cm/s. 2 Purge the column with carrier flow for at least 10 minutes before heating the oven.
  • Page 93: To Install A Packed Metal Column

    Maintaining the Purged Packed Inlet To Install a Packed Metal Column To Install a Packed Metal Column 1 Gather the following: • 7/16-inch, 9/16-inch, and 1/2-inch wrenches • Lint-free gloves 2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance > Maintenance Mode >...
  • Page 94 Maintaining the Purged Packed Inlet To Install a Packed Metal Column 13 Attach the column to the detector or detector adapter. Finger-tighten the nut. Adapter Ferrule Packed column 14 Using two wrenches, one on the adapter and one on the column nut, tighten the column nut an additional 1/4 turn (for a 1/8-inch column) or an additional 3/4 turn (for a 1/4-inch column).
  • Page 95: To Install A Packed Column Adapter Onto A Detector

    Maintaining the Purged Packed Inlet To Install a Packed Column Adapter onto a Detector To Install a Packed Column Adapter onto a Detector This procedure applies to the FID, FPD+, NPD, and TCD. 1 Gather the following: • Packed column adapter, see “Consumables and Parts for the Purged Packed Inlet”...
  • Page 96 Maintaining the Purged Packed Inlet To Install a Packed Column Adapter onto a Detector Detector fitting 5 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance > Maintenance Mode > Finished. After installation, heat the oven to 200 °C, cool it to a safe handling temperature, then re-tighten the column connection.
  • Page 97: To Install A Packed Glass Column

    Maintaining the Purged Packed Inlet To Install a Packed Glass Column To Install a Packed Glass Column 1 Gather the following: • 9/16-inch wrench • Two 1/4-inch brass nuts, see “Consumables and Parts for the Purged Packed Inlet” page 74. •...
  • Page 98 Maintaining the Purged Packed Inlet To Install a Packed Glass Column 8 Withdraw the column 1 to 2 mm from both the inlet and detector. Finger-tighten both column nuts. Overtightening the column nut or forcing it to bottom in either the inlet or detector may C AU T I O N shatter the column.
  • Page 99: To Condition A Packed Column

    Maintaining the Purged Packed Inlet To Condition a Packed Column To Condition a Packed Column 1 Gather the following: • Capillary adapter • Column nut and no-hole ferrule or blanking nut • Two 7/16-inch wrenches • 1/4-inch open-end wrench • Lint-free gloves Do not use hydrogen as the carrier for conditioning! It could vent into the oven and present an WAR N IN G...
  • Page 100: To Install Ferrules On A Packed Metal Column

    Maintaining the Purged Packed Inlet To Install Ferrules on a Packed Metal Column To Install Ferrules on a Packed Metal Column 1 Gather the following: • Wrenches • Stainless steel male Swagelok fitting, 1/4- or 1/8-inch od • Brass Swagelok nut and ferrule set, see “Consumables and Parts for the Purged Packed Inlet”...
  • Page 101: Maintaining The Coc Inlet

    Maintaining the COC Inlet Consumables and Parts for the Cool On-Column Inlet 102 Exploded Parts View of the Cool On-Column Inlet 105 To Install a Capillary Column with the Cool On-Column Inlet 106 To Check the Needle-to-Column Size on the COC Inlet 109 To Change a Septum on the COC Inlet 110 To Install an Insert on the COC Inlet 112 To Clean the COC Inlet 114...
  • Page 102: Consumables And Parts For The Cool On-Column Inlet

    Maintaining the COC Inlet Consumables and Parts for the Cool On-Column Inlet Consumables and Parts for the Cool On-Column Inlet See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 14 Recommended parts for injections onto 0.53-mm fused silica columns Column type Part number...
  • Page 103 Maintaining the COC Inlet Consumables and Parts for the Cool On-Column Inlet Table 16 Recommended parts for injections onto 0.32-mm fused silica columns (continued) Column type Part number Needle support assembly, 250/320 µm, for 7683B injector G2913-60978 Table 17 Recommended parts for injections onto 0.25-mm fused silica columns Column type Part number Insert, 0.25 mm id...
  • Page 104 Maintaining the COC Inlet Consumables and Parts for the Cool On-Column Inlet Table 19 Recommended septa for the COC inlet (continued) Description Part number/quantity For the duckbill septum Duckbill septum for manual injection only (must use cooling tower with duckbill) 19245-40050 (10/pk) Table 20 Nuts, ferrules, and hardware for capillary columns Column id (mm)
  • Page 105: Exploded Parts View Of The Cool On-Column Inlet

    Maintaining the COC Inlet Exploded Parts View of the Cool On-Column Inlet Exploded Parts View of the Cool On-Column Inlet Septum nut (0.53 mm) Cooling tower assembly Septum nut (0.25-mm and 0.32-mm columns) Duckbill septum Septum Spring Insert Ferrule Column nut Figure 26.
  • Page 106: To Install A Capillary Column With The Cool On-Column Inlet

    Maintaining the COC Inlet To Install a Capillary Column with the Cool On-Column Inlet To Install a Capillary Column with the Cool On-Column Inlet 1 Gather the following: • Column nut and ferrule, see “Consumables and Parts for the Cool On-Column Inlet” page 102.
  • Page 107 Maintaining the COC Inlet To Install a Capillary Column with the Cool On-Column Inlet 6 Break off the column end by supporting it against the column cutter opposite the scribe. Inspect the end with a magnifying loupe to make certain there are no burrs or jagged edges.
  • Page 108 Maintaining the COC Inlet To Install a Capillary Column with the Cool On-Column Inlet 12 Plug in the new column’s Smart ID key. See Figure Smart ID key Figure 27. Insert Smart ID key 13 Configure the new column. 14 Condition the column per the manufacturer’s recommendation. See To Condition a Capillary Column.
  • Page 109: To Check The Needle-To-Column Size On The Coc Inlet

    Maintaining the COC Inlet To Check the Needle-to-Column Size on the COC Inlet To Check the Needle-to-Column Size on the COC Inlet 1 Gather the following: • Insert, see “Consumables and Parts for the Cool On-Column Inlet” on page 102. •...
  • Page 110: To Change A Septum On The Coc Inlet

    Maintaining the COC Inlet To Change a Septum on the COC Inlet To Change a Septum on the COC Inlet 1 Gather the following: • Replacement septum, see “Consumables and Parts for the Cool On-Column Inlet” page 102. • Tweezers •...
  • Page 111 Maintaining the COC Inlet To Change a Septum on the COC Inlet For manual 200-µm injections with fused silica needle 4 Before making an injection, check the alignment of the entire assembly using the proper size syringe. 5 On the GC touchscreen select Maintenance > Inlets > Septum injections, then touch Reset Counter.
  • Page 112: To Install An Insert On The Coc Inlet

    Maintaining the COC Inlet To Install an Insert on the COC Inlet To Install an Insert on the COC Inlet 1 Gather the following: • Lint-free gloves • Replacement insert, see “Consumables and Parts for the Cool On-Column Inlet” page 102. 2 Place GC in maintenance mode: Maintenance >...
  • Page 113 Maintaining the COC Inlet To Install an Insert on the COC Inlet 7 Check that the insert is the correct size for both the needle and column. See “To Check the Needle-to-Column Size on the COC Inlet” on page 109. 8 Lower the new insert straight into the inlet from the top.
  • Page 114: To Clean The Coc Inlet

    Maintaining the COC Inlet To Clean the COC Inlet To Clean the COC Inlet 1 Gather the following: • 1/4-inch and 5/16-inch wrenches • Cleansing bath • Aqueous detergent • Distilled water • Methanol • Compressed, filtered, dry air or nitrogen •...
  • Page 115 Maintaining the COC Inlet To Clean the COC Inlet 8 Remove the spring and insert from the bath and rinse thoroughly with water and methanol. 9 Dry the spring and insert with compressed air or nitrogen. 10 Install the insert. See “To Install an Insert on the COC Inlet”...
  • Page 116: To Replace The 7693A Injector Needle Support Guide

    Maintaining the COC Inlet To Replace the 7693A Injector Needle Support Guide To Replace the 7693A Injector Needle Support Guide Before using a 7693A ALS to make injections onto the COC inlet, install the on-column needle support guide. Refer to the 7693A ALS documentation for complete details. Standard needle support guide On-column...
  • Page 117: To Replace A Needle In A Syringe

    Maintaining the COC Inlet To Replace a Needle in a Syringe To Replace a Needle in a Syringe 1 Gather the following, see Table 17, “Recommended parts for injections onto 0.25-mm fused silica columns,” on page 103: • Syringe barrel •...
  • Page 118: To Replace The Fused Silica Needle In A Syringe For The Coc Inlet

    Maintaining the COC Inlet To Replace the Fused Silica Needle in a Syringe for the COC Inlet To Replace the Fused Silica Needle in a Syringe for the COC Inlet The fused silica needle and syringe are only used with the cooling tower and duckbill septum NOTE for manual, on-column injections onto 200-µm columns.
  • Page 119: To Bakeout Contaminants From The Coc Inlet

    Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet To Bakeout Contaminants from the COC Inlet 1 Set the column flow to the normal operating value, or set the capillary column gas velocity to 30 cm/s. 2 Purge the column with carrier flow for at least 10 minutes before heating the oven. 3 Set the inlet mode to Oven Track.
  • Page 120 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet Maintaining Your GC...
  • Page 121: Maintaining The Mmi

    Maintaining the MMI Consumables and Parts for the Multimode Inlet 122 Exploded Parts View of the MMI 124 To Install the Column Nut Adapter 125 To Install a Capillary Column with the Multimode Inlet 126 To Change the Septum on the MMI 130 To Clean the Septum Seat in the Insert Assembly of the MMI 132 To Change the Liner and O-Ring on the MMI 134 To Replace the Filter in the Split Vent Trap for the MMI 137...
  • Page 122: Consumables And Parts For The Multimode Inlet

    Maintaining the MMI Consumables and Parts for the Multimode Inlet Consumables and Parts for the Multimode Inlet See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 21 Split, splitless, direct, and direct connect inlet liners Mode Description...
  • Page 123 Maintaining the MMI Consumables and Parts for the Multimode Inlet Table 22 Nuts, ferrules, and hardware for capillary columns (continued) Column id (mm) Description Typical use Part number/quantity .100 and .200 Ferrule, Vespel/graphite, 0.37-mm id 0.1-mm and 0.2-mm capillary columns 5062-3516 (10/pk) Ferrule, Vespel/graphite, 0.4-mm id 0.1-mm, 0.2-mm, and 0.25-mm capillary...
  • Page 124: Exploded Parts View Of The Mmi

    Maintaining the MMI Exploded Parts View of the MMI Exploded Parts View of the MMI Merlin cup Septum retainer nut Septum Merlin Microseal Insert assembly O-ring Split vent line Liner Cryo for CO2 Split vent line Inlet base Column nut Insulation Insulation cup Screws...
  • Page 125: To Install The Column Nut Adapter

    Maintaining the MMI To Install the Column Nut Adapter To Install the Column Nut Adapter 1 Gather the following, see “Consumables and Parts for the Multimode Inlet” on page 122: • Column nut adapter 2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance > Maintenance Mode >...
  • Page 126: To Install A Capillary Column With The Multimode Inlet

    Maintaining the MMI To Install a Capillary Column with the Multimode Inlet To Install a Capillary Column with the Multimode Inlet Do not use hydrogen as the carrier for conditioning! It could vent into the oven and present an WAR N IN G explosion hazard.
  • Page 127 Maintaining the MMI To Install a Capillary Column with the Multimode Inlet Ferrule Column nut Septum 6 Score the column using a glass scribing tool. The score must be square to ensure a clean break. 7 Break off the column end by supporting it against the column cutter opposite the scribe. Inspect the end with a magnifying loupe to make certain there are no burrs or jagged edges.
  • Page 128 Maintaining the MMI To Install a Capillary Column with the Multimode Inlet 10%12 mm 10 While holding the inlet base with a wrench thread the column nut into the inlet (but do not tighten). 11 Adjust the column position so that the septum contacts the bottom of the column nut. Finger-tighten the column nut until it begins to grip the column.
  • Page 129 Maintaining the MMI To Install a Capillary Column with the Multimode Inlet Smart ID key Figure 29. Insert Smart ID key 14 Configure the new column. 15 Condition the column per the manufacturer’s recommendation. See To Condition a Capillary Column. 16 Install the column into the detector.
  • Page 130: To Change The Septum On The Mmi

    Maintaining the MMI To Change the Septum on the MMI To Change the Septum on the MMI 1 Gather the following: • Replacement septum, see “Consumables and Parts for the Multimode Inlet” page 122. • Wrench, hex for changing septum •...
  • Page 131 Maintaining the MMI To Change the Septum on the MMI 1 mm 7 On the GC touchscreen, select Maintenance > Inlets > Septum injections, then touch Reset Counter. 8 The GC maintenance wizard will perform checks at the appropriate times, including Leak & Restriction tests, and will automatically reset the maintenance counters.
  • Page 132: To Clean The Septum Seat In The Insert Assembly Of The Mmi

    Maintaining the MMI To Clean the Septum Seat in the Insert Assembly of the MMI To Clean the Septum Seat in the Insert Assembly of the MMI 1 Gather the following: • Replacement septum, see “Consumables and Parts for the Multimode Inlet” page 122.
  • Page 133 Maintaining the MMI To Clean the Septum Seat in the Insert Assembly of the MMI 7 Use compressed air or nitrogen to blow away the pieces of steel wool and septum. 8 Line up the tab on the bottom of the insert assembly with the slot on the inlet body and push down to connect.
  • Page 134: To Change The Liner And O-Ring On The Mmi

    Maintaining the MMI To Change the Liner and O-Ring on the MMI To Change the Liner and O-Ring on the MMI 1 Gather the following: • Replacement O-ring, see “Consumables and Parts for the Multimode Inlet” page 122. • Replacement liner •...
  • Page 135 Maintaining the MMI To Change the Liner and O-Ring on the MMI 5 Grasp the liner with tweezers and pull it out. 6 Inspect the surface of the seal for contamination. If required, clean with cotton swabs. Maintaining Your GC...
  • Page 136 Maintaining the MMI To Change the Liner and O-Ring on the MMI Clean the inlet if there is visible or suspected contamination. See “To Clean the Multimode Inlet” on page 139. Clean O-ring residue from sealing surface. Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils. C AU T I O N 7 Slide a new O-ring onto the replacement liner.
  • Page 137: To Replace The Filter In The Split Vent Trap For The Mmi

    Maintaining the MMI To Replace the Filter in the Split Vent Trap for the MMI To Replace the Filter in the Split Vent Trap for the 1 Gather the following: • New split vent filter cartridge, see “Consumables and Parts for the Multimode Inlet” page 122.
  • Page 138 Maintaining the MMI To Replace the Filter in the Split Vent Trap for the MMI Filter and O-ring Knurled nut Figure 31. Slide trap assembly to expose the filter 6 Remove the old filter cartridge and two O-rings. 7 Verify the new O-rings are seated properly on the new filter cartridge. 8 Install the new filter cartridge then reassemble the trap using your fingers to fully tighten the knurled nut.
  • Page 139: To Clean The Multimode Inlet

    Maintaining the MMI To Clean the Multimode Inlet To Clean the Multimode Inlet This procedure explains how to clean the Multi Mode Inlet (MMI). Depending on the inlet mode used, the liner installed, and the cleanliness of the sample, the frequency of cleaning may range from weekly to monthly.
  • Page 140 Maintaining the MMI To Clean the Multimode Inlet 9 Using a second solvent moistened cotton swab, wipe the bottom of the inlet to remove any residue. Depending on the amount of use and maximum temperature used, the bottom of the inlet may still show some oxidation (light brown color). This is normal and will not affect inlet performance.
  • Page 141: To Bakeout Contaminants From The Mmi

    Maintaining the MMI To Bakeout Contaminants from the MMI To Bakeout Contaminants from the MMI 1 Put the inlet into split mode. 2 Set the column flow to the normal operating value, or set the capillary column gas velocity to 30 cm/s. 3 Set the inlet split vent flow to 200 mL/min.
  • Page 142 Maintaining the MMI To Bakeout Contaminants from the MMI Maintaining Your GC...
  • Page 143: Maintaining The Ptv Inlet

    Maintaining the PTV Inlet Consumables and Parts for the Programmable Temperature Vaporization Inlet 144 Exploded Parts View of the PTV Inlet 146 To Install a Capillary Column with the Programmable Temperature Vaporization Inlet 147 To Clean the Septumless Head on the PTV Inlet 150 To Replace the Septumless Head PTFE Ferrule on the PTV Inlet 152 To Change the Septum on the PTV Inlet 154 To Clean the Septum Seat in the Septum Head Assembly of the PTV Inlet 155...
  • Page 144: Consumables And Parts For The Programmable Temperature Vaporization Inlet

    Maintaining the PTV Inlet Consumables and Parts for the Programmable Temperature Vaporization Inlet Consumables and Parts for the Programmable Temperature Vaporization Inlet See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 24 PTV liners and ferrules Description Part number...
  • Page 145 Maintaining the PTV Inlet Consumables and Parts for the Programmable Temperature Vaporization Inlet Table 25 Other consumables and parts for the PTV inlet (continued) Description Part number Graphpak 2M inlet adapter, 0.53-mm id 5182-9762 Ferrules for Graphpak 2M inlet, 0.20-mm id 5182-9756 (10/pk) Ferrules for Graphpak 2M inlet, 0.25-mm id 5182-9768 (10/pk)
  • Page 146: Exploded Parts View Of The Ptv Inlet

    Maintaining the PTV Inlet Exploded Parts View of the PTV Inlet Exploded Parts View of the PTV Inlet Guide cap Septum nut Merlin cap PTFE ferrule Merlin Microseal Septum Septumless head Kalrez seal assembly Valve body Pressure spring Septum head Viton seal Carrier gas Sealing element...
  • Page 147: To Install A Capillary Column With The Programmable Temperature Vaporization Inlet

    Maintaining the PTV Inlet To Install a Capillary Column with the Programmable Temperature Vaporization Inlet To Install a Capillary Column with the Programmable Temperature Vaporization Inlet 1 Gather the following, see “Consumables and Parts for the Programmable Temperature Vaporization Inlet” on page 144: •...
  • Page 148 Maintaining the PTV Inlet To Install a Capillary Column with the Programmable Temperature Vaporization Inlet 7 Break off the column end by supporting it against the column cutter opposite the scribe. Inspect the end with a magnifying loupe to make certain that there are no burrs or jagged edges.
  • Page 149 Maintaining the PTV Inlet To Install a Capillary Column with the Programmable Temperature Vaporization Inlet Smart ID key Figure 33. Insert Smart ID key 13 Configure the new column. 14 Condition the column per the manufacturer’s recommendation. See To Condition a Capillary Column.
  • Page 150: To Clean The Septumless Head On The Ptv Inlet

    Maintaining the PTV Inlet To Clean the Septumless Head on the PTV Inlet To Clean the Septumless Head on the PTV Inlet 1 Gather the following: • Syringe with 23-gauge needle, see “Consumables and Parts for the Programmable Temperature Vaporization Inlet” on page 144.
  • Page 151 Maintaining the PTV Inlet To Clean the Septumless Head on the PTV Inlet Guide cap PTFE ferrule Ferrule, 1/16-inch Kalrez seal PTFE Valve body Carrier gas tube Pressure spring Viton seal Sealing element 6 Unscrew the guide cap from the head and remove the PTFE ferrule. 7 Carefully insert a syringe with a 23-gauge needle into the head to press the valve body and Kalrez seal slightly out of the head.
  • Page 152: To Replace The Septumless Head Ptfe Ferrule On The Ptv Inlet

    Maintaining the PTV Inlet To Replace the Septumless Head PTFE Ferrule on the PTV Inlet To Replace the Septumless Head PTFE Ferrule on the PTV Inlet 1 Gather the following: • Syringe with 23-gauge needle, see “Consumables and Parts for the Programmable Temperature Vaporization Inlet”...
  • Page 153 Maintaining the PTV Inlet To Replace the Septumless Head PTFE Ferrule on the PTV Inlet 12 Observe the inlet Total Flow. Tighten the guide cap until the Total Flow stops decreasing (typically near 8 mL/min). 13 Remove the syringe from the inlet and press 14 Restore the analytical method.
  • Page 154: To Change The Septum On The Ptv Inlet

    Maintaining the PTV Inlet To Change the Septum on the PTV Inlet To Change the Septum on the PTV Inlet 1 Gather the following: • Replacement septum, see “Consumables and Parts for the Programmable Temperature Vaporization Inlet” on page 144. •...
  • Page 155: To Clean The Septum Seat In The Septum Head Assembly Of The Ptv Inlet

    Maintaining the PTV Inlet To Clean the Septum Seat in the Septum Head Assembly of the PTV Inlet To Clean the Septum Seat in the Septum Head Assembly of the PTV Inlet 1 Gather the following: • Replacement septum, see “Consumables and Parts for the Programmable Temperature Vaporization Inlet”...
  • Page 156: To Change The Liner On The Ptv Inlet

    Maintaining the PTV Inlet To Change the Liner on the PTV Inlet To Change the Liner on the PTV Inlet 1 Gather the following: • Installation tool for 3D ferrules, see “Consumables and Parts for the Programmable Temperature Vaporization Inlet” on page 144.
  • Page 157 Maintaining the PTV Inlet To Change the Liner on the PTV Inlet Ferrule guide Graphpak 3D ferrule Compression fitting Open baffle liner 6 Slide the compression fitting onto the longer, straight end of the new liner with the threads pointing toward the end of the liner. 7 Place a Graphpak 3D ferrule on the same end of the liner with the recessed graphite end towards the compression fitting.
  • Page 158: To Replace The Inlet Adapter For The Ptv Inlet

    Maintaining the PTV Inlet To Replace the Inlet Adapter for the PTV Inlet To Replace the Inlet Adapter for the PTV Inlet 1 Select from the following list an adapter with the smallest hole diameter that will accept the column. The adapter number is stamped on the side of the adapter. Also see “Consumables and Parts for the Programmable Temperature Vaporization Inlet”...
  • Page 159 Maintaining the PTV Inlet To Replace the Inlet Adapter for the PTV Inlet Silver seal Graphpak inlet adapter Column nut Ferrule for Graphpak adapter 5 Remove the inlet adapter and discard the old silver seal. 6 Insert a new silver seal into the adapter and finger-tighten the adapter onto the inlet. Tighten an additional 1/16 to 1/8 turn with a wrench;...
  • Page 160: To Replace The Filter In The Split Vent Trap For The Ptv Inlet

    Maintaining the PTV Inlet To Replace the Filter in the Split Vent Trap for the PTV Inlet To Replace the Filter in the Split Vent Trap for the PTV Inlet 1 Gather the following: • New split vent filter cartridge, see “Consumables and Parts for the Programmable Temperature Vaporization Inlet”...
  • Page 161 Maintaining the PTV Inlet To Replace the Filter in the Split Vent Trap for the PTV Inlet Filter and O-ring Knurled nut Figure 35. Slide trap assembly to expose the filter 6 Remove the old filter cartridge and two O-rings. 7 Verify the new O-rings are seated properly on the new filter cartridge.
  • Page 162: To Bakeout Contaminants From The Ptv Inlet

    Maintaining the PTV Inlet To Bakeout Contaminants from the PTV Inlet To Bakeout Contaminants from the PTV Inlet 1 Put the inlet into split mode. 2 Set the column flow to the normal operating value, or set the capillary column gas velocity to 30 cm/s.
  • Page 163: Maintaining The

    Maintaining the VI Consumables and Parts for the Volatiles Interface 164 Exploded Parts View of the VI 166 To Install a Capillary Column with the Volatiles Interface 167 To Remove the VI Interface 171 To Clean the VI 173 To Install the VI Interface 174 To Replace the Filter in the Split Vent Trap for the VI 175 To Bakeout Contaminants from the VI Inlet 177 Maintaining Your GC...
  • Page 164: Consumables And Parts For The Volatiles Interface

    10 Maintaining the VI Consumables and Parts for the Volatiles Interface Consumables and Parts for the Volatiles Interface See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 27 Parts for the VI Description Part number...
  • Page 165 10 Maintaining the VI Consumables and Parts for the Volatiles Interface Table 28 Nuts, ferrules, and hardware for capillary columns (continued) Column id (mm) Description Typical use Part number/quantity Ferrule, no-hole Testing 5181-3308 (10/pk) Capillary column blanking nut Testing–use with any ferrule 5020-8294 Column nut, universal Connect column to inlet or detector...
  • Page 166: Exploded Parts View Of The Vi

    10 Maintaining the VI Exploded Parts View of the VI Exploded Parts View of the VI Ferrule Clamping plate Volatiles interface Ferrule, column Column nut, long Figure 36. Volatiles interface exploded parts Maintaining Your GC...
  • Page 167: To Install A Capillary Column With The Volatiles Interface

    10 Maintaining the VI To Install a Capillary Column with the Volatiles Interface To Install a Capillary Column with the Volatiles Interface 1 Gather the following: • Long (65 mm) column nut • Column • Ferrule • Column cutter • Isopropanol •...
  • Page 168 10 Maintaining the VI To Install a Capillary Column with the Volatiles Interface Ferrule Column nut Septum 5 Score the column using a glass scribing tool. The score must be square to ensure a clean break. 6 Break off the column end by supporting it against the column cutter opposite the scribe. Inspect the end with a magnifying loupe to make certain that there are no burrs or jagged edges.
  • Page 169 10 Maintaining the VI To Install a Capillary Column with the Volatiles Interface 9 Insert the column into the interface and finger-tighten the column nut. 10 Adjust the column (not the septum) position until the septum is snug against the bottom of the nut.
  • Page 170 10 Maintaining the VI To Install a Capillary Column with the Volatiles Interface 20 After the GC becomes ready, wait 10 minutes then ignite the detector flame or adjust offset on the NPD bead. Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven, WAR N IN G inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
  • Page 171: To Remove The Vi Interface

    10 Maintaining the VI To Remove the VI Interface To Remove the VI Interface 1 Gather the following: • 1/4-inch and 7-mm wrench • T-20 Torx screwdriver 2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance > Maintenance Mode >...
  • Page 172 10 Maintaining the VI To Remove the VI Interface 8 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance > Maintenance Mode > Finished. Maintaining Your GC...
  • Page 173: To Clean The Vi

    10 Maintaining the VI To Clean the VI To Clean the VI 1 Gather the following: • 1/4-inch and 7-mm wrench • T-20 Torx screwdriver • Lint-free gloves 2 Manually set the inlet and oven temperature to < 40 °C, and wait for the inlet, oven, and other parts you might come into contact with inside the oven, to cool before continuing.
  • Page 174: To Install The Vi Interface

    10 Maintaining the VI To Install the VI Interface To Install the VI Interface 1 Gather the following: • 1/4-inch and 7-mm wrench • T-20 Torx screwdriver 2 Attach the split vent line and pressure sensing lines and finger-tighten the nuts. Tighten an additional 1/4 turn with a wrench.
  • Page 175: To Replace The Filter In The Split Vent Trap For The Vi

    10 Maintaining the VI To Replace the Filter in the Split Vent Trap for the VI To Replace the Filter in the Split Vent Trap for the VI 1 Gather the following: • New split vent filter cartridge, see “Consumables and Parts for the Volatiles Interface”...
  • Page 176 10 Maintaining the VI To Replace the Filter in the Split Vent Trap for the VI Filter and O-ring Knurled nut Figure 39. Slide trap assembly to expose the filter 6 Remove the old filter cartridge and two O-rings. 7 Verify the new O-rings are seated properly on the new filter cartridge. 8 Install the new filter cartridge then reassemble the trap using your fingers to fully tighten the knurled nut.
  • Page 177: To Bakeout Contaminants From The Vi Inlet

    10 Maintaining the VI To Bakeout Contaminants from the VI Inlet To Bakeout Contaminants from the VI Inlet 1 Put the inlet into split mode. 2 Set the column flow to the normal operating value, or set the capillary column gas velocity to 30 cm/s.
  • Page 178 10 Maintaining the VI To Bakeout Contaminants from the VI Inlet Maintaining Your GC...
  • Page 179: Maintaining The Fid

    Maintaining the FID Consumables and Parts for the FID 180 Exploded Parts Views of the Flame Ionization Detector 182 To Install a Capillary Column in the FID 184 To Replace an FID Jet 187 To Perform Maintenance on the FID Collector Assembly 190 To Check the FID Leakage Current 197 To Check the FID Baseline 198 To Install the Optional FID PTFE Chimney Insert 199...
  • Page 180: Consumables And Parts For The Fid

    11 Maintaining the FID Consumables and Parts for the FID Consumables and Parts for the FID See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 29 Nuts, ferrules, and hardware for capillary columns Column id (mm) Description...
  • Page 181 11 Maintaining the FID Consumables and Parts for the FID Table 30 FID parts and subassemblies Description Part number/quantity Screw, M4 × 25 mm, Torx, T20 0515-2712 (3/pk) PTFE chimney (optional) 19231-21050 Collector assembly G4591-60691 1/8-inch Packed column adapter assembly G3435-81340 1/4-inch Packed column adapter assembly G3435-81330...
  • Page 182: Exploded Parts Views Of The Flame Ionization Detector

    11 Maintaining the FID Exploded Parts Views of the Flame Ionization Detector Exploded Parts Views of the Flame Ionization Detector PTFE chimney 25-mm screws Collector assembly Gasket Spring, FID interconnect Mounting pallet Ferrule Column adapter Ferrule Column nut Figure 40. Flame ionization detector (FID) assembly exploded parts Maintaining Your GC...
  • Page 183 11 Maintaining the FID Exploded Parts Views of the Flame Ionization Detector 25-mm screws Collector assembly Gasket Figure 41. FID exploded parts view Collector nut Spring washer 25-mm screws Ignitor (glow plug) Ignitor castle assembly with O-ring Collector Upper collector assembly insulator Collector body...
  • Page 184: To Install A Capillary Column In The Fid

    11 Maintaining the FID To Install a Capillary Column in the FID To Install a Capillary Column in the FID 1 Gather the following materials, see “Consumables and Parts for the FID” on page 180: • Column • Ferrule(s) • Column nut •...
  • Page 185 11 Maintaining the FID To Install a Capillary Column in the FID 5 Break off the column end by supporting it against the column cutter opposite the scribe. Inspect the end with a magnifying loupe to make certain there are no burrs or jagged edges.
  • Page 186 11 Maintaining the FID To Install a Capillary Column in the FID If the column id is 0.1 mm or less position the column so it extends above the ferrule by 48 mm. Slide the septum up to hold the column nut and ferrule at this fixed position. 48 mm c Insert the column into the detector.
  • Page 187: To Replace An Fid Jet

    11 Maintaining the FID To Replace an FID Jet To Replace an FID Jet 1 Gather the following: • Replacement jet • T-20 Torx screwdriver • 1/4-inch nut driver • Tweezers • Compressed, filtered, dry air or nitrogen • Solvent that will clean the type of deposits in your detector •...
  • Page 188 11 Maintaining the FID To Replace an FID Jet Handle the clean or new jet only with tweezers, or wear gloves. C AU T I O N 6 Loosen the jet, then lift it out of the housing with tweezers. Interconnector spring 7 Clean the detector base cavity using solvent, a swab, and compressed air or nitrogen.
  • Page 189 11 Maintaining the FID To Replace an FID Jet 10 Install the castle assembly. 11 Select Finished, then select OK to exit the GC maintenance wizard. 12 Reset the EMF counter. 13 Attach the capillary column to the detector. a Install the column in the detector. See “To Install a Capillary Column in the FID”...
  • Page 190: To Perform Maintenance On The Fid Collector Assembly

    11 Maintaining the FID To Perform Maintenance on the FID Collector Assembly To Perform Maintenance on the FID Collector Assembly Perform only the steps and gather only the parts that apply to the desired maintenance NOTE task(s). 1 Gather the following: •...
  • Page 191 11 Maintaining the FID To Perform Maintenance on the FID Collector Assembly c Turn the nut counterclockwise by hand. Remove the ignitor and copper washer. 4 If replacing only the FID ignitor assembly with copper washer, skip to step 16 for assembly.
  • Page 192 11 Maintaining the FID To Perform Maintenance on the FID Collector Assembly 7 Remove the gasket from the bottom of the assembly, if necessary. 8 Remove the FID ignitor castle. a Loosen the collector nut. b Remove the collector nut and the spring washer. c Lift the castle out of the collector housing.
  • Page 193 11 Maintaining the FID To Perform Maintenance on the FID Collector Assembly b Remove the lower insulator with tweezers and place the parts on a clean cloth. 11 Remove the collector housing from the mount, if necessary. 12 Use tweezers to remove the gasket from the bottom of the housing. The collector assembly is now completely disassembled.
  • Page 194 11 Maintaining the FID To Perform Maintenance on the FID Collector Assembly 14 Install the collector insulators. a Insert one of the insulators into the base of the housing. Seat the insulator with the flat surface facing out of the housing. b Insert the long end of collector into the housing and lower insulator.
  • Page 195 11 Maintaining the FID To Perform Maintenance on the FID Collector Assembly b Insert the FID castle into the collector housing. c Install the spring washer over the castle. d Install the collector nut over the castle and tighten firmly. The seal should be airtight. Maintain the orientation of the ignitor hole with the base as shown below.
  • Page 196 11 Maintaining the FID To Perform Maintenance on the FID Collector Assembly 17 Lower the collector assembly into the housing. 18 Insert the three screws and tighten (to 18-inch-pounds). 19 Connect the ignitor extension cable. 20 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance > Maintenance Mode >...
  • Page 197: To Check The Fid Leakage Current

    11 Maintaining the FID To Check the FID Leakage Current To Check the FID Leakage Current 1 Load the analytical method. • Make sure flows are acceptable for ignition. • Heat the detector to operating temperature or 300 °C. 2 Perform a Leakage Current Test from the touchscreen: Diagnostics > Diagnostics Tests > Detectors, select FID# Leakage Current Tests.
  • Page 198: To Check The Fid Baseline

    11 Maintaining the FID To Check the FID Baseline To Check the FID Baseline 1 With the column installed, load your checkout method. 2 Set the oven temperature to 35 °C. 3 Show the detector output signal in Status. On the GC touchscreen go to Home > Status listing >...
  • Page 199: To Install The Optional Fid Ptfe Chimney Insert

    11 Maintaining the FID To Install the Optional FID PTFE Chimney Insert To Install the Optional FID PTFE Chimney Insert Be careful! The detector may be hot enough to cause burns. If the detector is hot, wear WAR N IN G heat-resistant gloves to protect your hands.
  • Page 200: To Bakeout The Fid

    11 Maintaining the FID To Bakeout the FID To Bakeout the FID 1 Bakeout the FID with the column installed or uninstalled. If uninstalled, gather the following, “Consumables and Parts for the FID” on page 180: • Capillary adapter (adaptable FID only) •...
  • Page 201: Maintaining The Tcd

    Maintaining the TCD Consumables and Parts for the Thermal Conductivity Detector 202 To Install a Capillary Column in the TCD 204 To Bakeout Contaminants from the TCD 207 Maintaining Your GC...
  • Page 202: Consumables And Parts For The Thermal Conductivity Detector

    12 Maintaining the TCD Consumables and Parts for the Thermal Conductivity Detector Consumables and Parts for the Thermal Conductivity Detector See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Standard TCD column hardware Table 32 Nuts, ferrules, and hardware for capillary columns Column id (mm)
  • Page 203 12 Maintaining the TCD Standard TCD column hardware Table 33 Packed column adapters Description Part number/quantity 1/8-inch Packed column adapter assembly G3435-81340 1/4-inch Packed column adapter assembly G3435-81330 Maintaining Your GC...
  • Page 204: To Install A Capillary Column In The Tcd

    12 Maintaining the TCD To Install a Capillary Column in the TCD To Install a Capillary Column in the TCD 1 Gather the following materials, see “Consumables and Parts for the Thermal Conductivity Detector” on page 202: • Column • Ferrule(s) •...
  • Page 205 12 Maintaining the TCD To Install a Capillary Column in the TCD 5 Break off the column end by supporting it against the column cutter opposite the scribe. Inspect the end with a magnifying loupe to make certain there are no burrs or jagged edges.
  • Page 206 12 Maintaining the TCD To Install a Capillary Column in the TCD If the column id is 0.1 mm or less position the column so it extends above the ferrule by 48 mm (capillary optimized fitting). Slide the septum up to hold the column nut and ferrule at this fixed position.
  • Page 207: To Bakeout Contaminants From The Tcd

    12 Maintaining the TCD To Bakeout Contaminants from the TCD To Bakeout Contaminants from the TCD The bakeout can be performed with the column installed or the detector capped. If the column is not installed, you must turn off the TCD filament and cap the detector column C AU T I O N fitting to prevent irreparable damage to the filament caused by oxygen entering the detector.
  • Page 208 12 Maintaining the TCD To Bakeout Contaminants from the TCD Maintaining Your GC...
  • Page 209: Maintaining The Npd

    Maintaining the NPD Consumables and Parts for the NPD 210 Exploded Parts View of the Nitrogen-Phosphorus Detector 212 To Install a Capillary Column in the NPD 213 To Replace the NPD Bead Assembly 216 To Maintain the NPD Collector, Ceramic Insulators, and Jet 220 To Check the NPD Leakage Current 225 To Bakeout the NPD 226 Maintaining Your GC...
  • Page 210: Consumables And Parts For The Npd

    13 Maintaining the NPD Consumables and Parts for the NPD Consumables and Parts for the NPD See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 34 NPD parts Description Part number/quantity...
  • Page 211 13 Maintaining the NPD Consumables and Parts for the NPD Table 35 Nuts, ferrules, and hardware for capillary columns (continued) Column id (mm) Description Typical use Part number/quantity .250 Ferrule, Vespel/graphite, 0.4-mm id 0.1-mm, 0.2-mm, and 0.25-mm capillary 5181-3323 (10/pk) columns Ferrule, graphite, 0.5-mm id 0.1-mm, 0.2-mm, 0.25-mm, and 0.32-mm...
  • Page 212: Exploded Parts View Of The Nitrogen-Phosphorus Detector

    13 Maintaining the NPD Exploded Parts View of the Nitrogen-Phosphorus Detector Exploded Parts View of the Nitrogen-Phosphorus Detector NPD cover Screws Bead assembly Screws Metal C-ring, top Ceramic insulator, upper Collector Metal C-ring, bottom Ceramic insulator, lower Metal C-ring, bottom Electrometer Insulation cover Column nut...
  • Page 213: To Install A Capillary Column In The Npd

    13 Maintaining the NPD To Install a Capillary Column in the NPD To Install a Capillary Column in the NPD 1 Gather the following materials: • Column • Ferrule(s), see “Consumables and Parts for the NPD” on page 210. • Column nut •...
  • Page 214 13 Maintaining the NPD To Install a Capillary Column in the NPD 5 Break off the column end by supporting it against the column cutter opposite the scribe. Inspect the end with a magnifying loupe to make certain there are no burrs or jagged edges.
  • Page 215 13 Maintaining the NPD To Install a Capillary Column in the NPD If the column id is 0.1 mm or less position the column so it extends above the ferrule by 48 mm (capillary optimized fitting). Slide the septum up to hold the column nut and ferrule at this fixed position.
  • Page 216: To Replace The Npd Bead Assembly

    13 Maintaining the NPD To Replace the NPD Bead Assembly To Replace the NPD Bead Assembly 1 Gather the following: • Replacement NPD bead assembly, see “Consumables and Parts for the NPD” page 210. • Lint-free gloves • T-10 Torx screwdriver 2 Set GC options for a new bead.
  • Page 217 13 Maintaining the NPD To Replace the NPD Bead Assembly Figure 44. Remove NPD cover 9 Twist the ring to disconnect the bead assembly cable. Push and twist the lock so that the button slides up in the groove, then pull the cable ends apart. Maintaining Your GC...
  • Page 218 13 Maintaining the NPD To Replace the NPD Bead Assembly 10 Remove the 3 T-10 Torx screws from the bead assembly. 11 Gently lift up and remove the old bead assembly. Avoid bumping the bead on the sides of the collector. 12 Remove the protective cap covering the new bead.
  • Page 219 13 Maintaining the NPD To Replace the NPD Bead Assembly 17 Close the NPD cover, install the GC detector top cover, and install the electronics top cover. All covers must be closed to get a stable NPD baseline. 18 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance > Maintenance Mode >...
  • Page 220: To Maintain The Npd Collector, Ceramic Insulators, And Jet

    13 Maintaining the NPD To Maintain the NPD Collector, Ceramic Insulators, and Jet To Maintain the NPD Collector, Ceramic Insulators, and Jet When replacing the jet, always install a new collector, ceramic insulators, and metal C-rings. When replacing the collector, Agilent recommends replacing the ceramic insulators and metal C-rings.
  • Page 221 13 Maintaining the NPD To Maintain the NPD Collector, Ceramic Insulators, and Jet This step exposes the interconnect spring. Be careful not to touch or disfigure the spring C AU T I O N while working on the FID. Any dirt or bending will reduce the sensitivity of your detector. Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
  • Page 222 13 Maintaining the NPD To Maintain the NPD Collector, Ceramic Insulators, and Jet 8 Remove the screws that secure the interconnect cover, then lift up the cover and set aside. 9 Remove the screws that secure the electrometer, then lift up and set aside. 10 Pull the electrometer away from the detector to free the interconnect.
  • Page 223 13 Maintaining the NPD To Maintain the NPD Collector, Ceramic Insulators, and Jet Metal C-ring, top Ceramic insulator, upper Collector Figure 45. NPD collector, upper insulator, and metal C-ring 13 Use tweezers to remove the lower ceramic insulator and the two small metal C-rings located above and below the collector.
  • Page 224 13 Maintaining the NPD To Maintain the NPD Collector, Ceramic Insulators, and Jet 17 Pull the jet straight out of the detector. Use tweezers, if necessary. The adaptable NPD jet is longer than the capillary optimized NPD extended jet and should C AU T I O N never be installed in a capillary optimized detector.
  • Page 225: To Check The Npd Leakage Current

    13 Maintaining the NPD To Check the NPD Leakage Current To Check the NPD Leakage Current 1 Load the analytical method. 2 Turn Off the Bead. • Leave the NPD at operating temperature • Leave flows on or off 3 Show the detector output signal in Status. 4 Verify that the output (leakage current) is stable and <...
  • Page 226: To Bakeout The Npd

    13 Maintaining the NPD To Bakeout the NPD To Bakeout the NPD If using hydrogen as a carrier gas, turn off the hydrogen supply and cap the end of the column WAR N IN G to prevent an oven explosion. 1 Bakeout the NPD with the column installed or uninstalled.
  • Page 227: Maintaining The Ecd

    Maintaining the ECD Important Safety Information About the ECD 228 Consumables and Parts for the ECD 230 Exploded Parts View of the Electron Capture Detector 232 To Replace the ECD Fused Silica Indented Mixing Liner and Install the Makeup Gas Adapter 233 To Install a Capillary Column in the ECD 236 To Bakeout the ECD 238...
  • Page 228: Important Safety Information About The Ecd

    14 Maintaining the ECD Important Safety Information About the ECD Important Safety Information About the ECD The ECD contains a cell plated with Ni, a radioactive isotope. The beta particles released at the energy level in the detector have little penetrating power—the surface layer of the skin or a few sheets of paper will stop most of them—but they may be hazardous if the isotope is ingested or inhaled.
  • Page 229 10 CFR Part 20 (including Appendix B), or the applicable state regulation. For other countries, consult with the appropriate agency for equivalent requirements. Agilent Technologies recommends a vent line internal diameter of 6 mm (1/4-inch) or greater. With a line of this diameter, the length is not critical.
  • Page 230: Consumables And Parts For The Ecd

    14 Maintaining the ECD Consumables and Parts for the ECD Consumables and Parts for the ECD See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 36 ECD consumables and parts Description Part number/quantity...
  • Page 231 14 Maintaining the ECD Consumables and Parts for the ECD Table 37 Nuts, ferrules, and hardware for capillary columns (continued) Column id (mm) Description Typical use Part number/quantity Ferrule, graphite, 0.4-mm id 500-2114 (10/pk) Column nut, finger-tight (for .100- to Connect column to inlet or detector 5020-8292 .320-mm columns)
  • Page 232: Exploded Parts View Of The Electron Capture Detector

    14 Maintaining the ECD Exploded Parts View of the Electron Capture Detector Exploded Parts View of the Electron Capture Detector Perforated cover Warning tag Electrometer Ferrules Adapter nut Liner Makeup gas line Makeup gas adapter Column nut Figure 47. Electron Capture Detector (ECD) exploded parts Maintaining Your GC...
  • Page 233: To Replace The Ecd Fused Silica Indented Mixing Liner And Install The Makeup Gas Adapter

    14 Maintaining the ECD To Replace the ECD Fused Silica Indented Mixing Liner and Install the Makeup Gas Adapter To Replace the ECD Fused Silica Indented Mixing Liner and Install the Makeup Gas Adapter 1 Gather the following: • Fused silica indented mixing liner, see “Consumables and Parts for the ECD”...
  • Page 234 14 Maintaining the ECD To Replace the ECD Fused Silica Indented Mixing Liner and Install the Makeup Gas Adapter 5 Unscrew and remove the adapter cap. Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils. C AU T I O N 6 Remove the fused silica indented mixing liner and inspect.
  • Page 235 14 Maintaining the ECD To Replace the ECD Fused Silica Indented Mixing Liner and Install the Makeup Gas Adapter 10 Slowly install the adapter straight into the detector fitting. Jiggle the adapter, if necessary, to make sure it is seated all the way into the detector fitting. Be careful not to break the column end.
  • Page 236: To Install A Capillary Column In The Ecd

    14 Maintaining the ECD To Install a Capillary Column in the ECD To Install a Capillary Column in the ECD 1 Gather the following: • Ferrule, see “Consumables and Parts for the ECD” on page 230. • Column nut • Septum •...
  • Page 237 14 Maintaining the ECD To Install a Capillary Column in the ECD 6 Break off the column end by supporting it against the column cutter opposite the scribe. Inspect the end with a magnifying loupe to make certain there are no burrs or jagged edges.
  • Page 238: To Bakeout The Ecd

    14 Maintaining the ECD To Bakeout the ECD To Bakeout the ECD Detector disassembly and/or cleaning procedures other than thermal should be performed WAR N IN G only by personnel trained and licensed appropriately to handle radioactive materials. Trace amounts of radioactive Ni may be removed during other procedures, causing possible hazardous exposure to b- and x-radiation.
  • Page 239 14 Maintaining the ECD To Bakeout the ECD 8 Allow thermal cleaning to continue for several hours and then cool the system to normal operating temperatures. The figure below shows detector output during a typical cleaning cycle. 9 Check the ECD output value on the control table. It should be lower than the first reading. If it is not, contact your Agilent service representative.
  • Page 240 14 Maintaining the ECD To Bakeout the ECD Maintaining Your GC...
  • Page 241: Maintaining The Fpd

    Maintaining the FPD+ Consumables and Parts for the FPD+ 242 Exploded Parts View of the Flame Photometric Detector Plus 244 To Install a Capillary Column to the FPD+ 245 To Change the FPD+ Wavelength Filter 247 To Remove the FPD+ Cover 250 To Replace the FPD+ Ignitor 251 To Install the FPD+ Cover 253 Cleaning the FPD+ Brazement 254...
  • Page 242: Consumables And Parts For The Fpd

    15 Maintaining the FPD+ Consumables and Parts for the FPD+ Consumables and Parts for the FPD+ See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 38 FPD supplies Description Part number/quantity...
  • Page 243 15 Maintaining the FPD+ Consumables and Parts for the FPD+ Table 39 Nuts, ferrules, and hardware for capillary columns (continued) Column id (mm) Description Typical use Part number/quantity .320 Ferrule, Vespel/graphite, 0.5-mm id 0.32-mm capillary columns 5062-3514 (10/pk) Ferrule, graphite, 0.5-mm id 0.1-mm, 0.2-mm, 0.25-mm, and 0.32-mm 5080-8853 (10/pk) capillary columns...
  • Page 244: Exploded Parts View Of The Flame Photometric Detector Plus

    15 Maintaining the FPD+ Exploded Parts View of the Flame Photometric Detector Plus Exploded Parts View of the Flame Photometric Detector Plus Single wavelength cover Filter Spacer (for sulfur filter) Ignitor (glow plug) Spring Collar Screw PMT assembly Ignitor cable assembly Column ferrule Adapter ferrule...
  • Page 245: To Install A Capillary Column To The Fpd

    15 Maintaining the FPD+ To Install a Capillary Column to the FPD+ To Install a Capillary Column to the FPD+ 1 Gather the following: • Column measuring tool, see “Consumables and Parts for the FPD+” on page 242. • Column cutter •...
  • Page 246 15 Maintaining the FPD+ To Install a Capillary Column to the FPD+ Good 8 Remove the column, nut, and swaged ferrule from the tool. 9 Wipe the column walls with a tissue dampened with isopropanol to remove fingerprints and dust. 10 Carefully thread the swaged column up into the detector fitting.
  • Page 247: To Change The Fpd+ Wavelength Filter

    15 Maintaining the FPD+ To Change the FPD+ Wavelength Filter To Change the FPD+ Wavelength Filter Do not touch the filter with your bare hands. For optimum performance and to avoid C AU T I O N scratches, use lint-free gloves for assembling and inserting the filter into the assembly. 1 Gather the following: •...
  • Page 248 15 Maintaining the FPD+ To Change the FPD+ Wavelength Filter 5 To prevent light from damaging the PMT, cap the end or place it face down. 6 Place a clean cloth under the filter housing to catch the filter. • For phosphorus filter, use the sharpened wooden tip of a toothpick or cotton swab to dislodge the filter from the housing.
  • Page 249 15 Maintaining the FPD+ To Change the FPD+ Wavelength Filter Sulfur filter Filter spacer Do not use cleaning fluids. Cleaning fluids will damage lens coatings. C AU T I O N 7 Clean the new filter with lens tissue. Filters are designed for the light of the flame to pass through in a specific direction. The C AU T I O N triangle (on the edge of the phosphorus filter) and the arrow (on the edge of the sulfur filter) should face away from the flame and toward the PMT.
  • Page 250: To Remove The Fpd+ Cover

    15 Maintaining the FPD+ To Remove the FPD+ Cover To Remove the FPD+ Cover 1 Gather the following: • T-20 Torx screwdriver 2 Turn off the flame, then turn off the GC. When turning off the GC, first turn off the flame to prevent condensation from dripping into C AU T I O N the jet and column.
  • Page 251: To Replace The Fpd+ Ignitor

    15 Maintaining the FPD+ To Replace the FPD+ Ignitor To Replace the FPD+ Ignitor 1 Gather the following: • Ignitor replacement kit, see “Consumables and Parts for the FPD+” on page 242. • Torx screwdrivers, T-20 and T-10 • 5/16-inch nut driver (or wrench) 2 Place GC in maintenance mode: Maintenance >...
  • Page 252 15 Maintaining the FPD+ To Replace the FPD+ Ignitor 6 Install the new ignitor assembly and tighten with the nut driver. Do not overtighten. (If the ignitor comes with a copper washer, discard the copper washer.) 7 Replace the ignitor collet and cable assembly and tighten the screw. Align the collet set screw as shown.
  • Page 253: To Install The Fpd+ Cover

    15 Maintaining the FPD+ To Install the FPD+ Cover To Install the FPD+ Cover 1 Gather the following: • T-20 Torx screwdriver 2 Install the cover. Single-wavelength detector: a Start the two screws on the right side of the cover. b Start and tighten the screws at the base on the left side.
  • Page 254: Cleaning The Fpd+ Brazement

    15 Maintaining the FPD+ Cleaning the FPD+ Brazement Cleaning the FPD+ Brazement The brazement uses an inert coating layer. Abrasives may scratch this layer. Hard scrubbing C AU T I O N can scratch this layer. Solutions or soaps with a pH > 8 can also damage this layer. Do not steam clean.
  • Page 255: Maintaining Epc Modules

    Maintaining EPC Modules Consumables and Parts for the Aux EPC 256 Installing or Replacing Frits in the Aux EPC 258 Consumables and Parts for the PCM 260 Calibrating the PCM Interface 261 Installing or Replacing Frits in the PCM 262 This chapter describes the maintenance procedures for Auxiliary EPC (Aux) and Pneumatics Control Module (PCM).
  • Page 256: Consumables And Parts For The Aux Epc

    16 Maintaining EPC Modules Consumables and Parts for the Aux EPC Consumables and Parts for the Aux EPC See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 40 Aux EPC consumables Description Part number...
  • Page 257 16 Maintaining EPC Modules Consumables and Parts for the Aux EPC Table 41 Auxiliary channel frits (continued) Frit marking Flow resistance Flow characteristic Often used with None (brass tube) Zero No restriction Headspace vial pressurization, purged splitter and Deans Switch when using backflush Maintaining Your GC...
  • Page 258: Installing Or Replacing Frits In The Aux Epc

    16 Maintaining EPC Modules Installing or Replacing Frits in the Aux EPC Installing or Replacing Frits in the Aux EPC To install or replace a frit in the Aux EPC block: 1 Gather the following: • T-10 Torx screwdriver • Appropriate O-ring and frit •...
  • Page 259 16 Maintaining EPC Modules Installing or Replacing Frits in the Aux EPC Screw Molded Partner fitting Frit O-ring Tubing weldment Figure 50. Replace an Aux EPC frit (2 flow tube channels not shown) 10 Place the tubing weldment into the Aux EPC module, and secure in place using the molded partner fitting and screw.
  • Page 260: Consumables And Parts For The Pcm

    16 Maintaining EPC Modules Consumables and Parts for the PCM Consumables and Parts for the PCM See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 42 PCM consumables Description Part number...
  • Page 261: Calibrating The Pcm Interface

    16 Maintaining EPC Modules Calibrating the PCM Interface Calibrating the PCM Interface The interface's flow module contains a pressure sensor that must be zeroed after it is installed on the GC. Calibration ensures an accurate interface pressure display. Do not connect the carrier gas to the flow module until you have zeroed the interface's pressure sensor.
  • Page 262: Installing Or Replacing Frits In The Pcm

    16 Maintaining EPC Modules Installing or Replacing Frits in the PCM Installing or Replacing Frits in the PCM To install or replace a frit: 1 Gather the following: • O-rings, package of 12, p/n 5180-4181 • Frit • Tweezers 2 Turn off the gas supply to the channel. 3 Select the appropriate frit.
  • Page 263 16 Maintaining EPC Modules Installing or Replacing Frits in the PCM When hydrogen is used, dangerously high flows are possible if insufficient flow resistance is WAR N IN G provided downstream of the supply tube. Always use either the High (Blue dot) or Medium (Red dot) frit with hydrogen.
  • Page 264 16 Maintaining EPC Modules Installing or Replacing Frits in the PCM Maintaining Your GC...
  • Page 265: Maintaining A Valve

    Maintaining a Valve Consumables and Parts for Valves 266 Exploded Parts View of GC Rotary Valves 267 To Replace a Gas Sampling Valve Loop 268 To Align a Rotary Valve Rotor 270 To Replace a Rotary Valve in the Valve Box 271 To Remove the Upper Valve Box 274 To Install the Upper Valve Box 275 Maintaining Your GC...
  • Page 266: Consumables And Parts For Valves

    17 Maintaining a Valve Consumables and Parts for Valves Consumables and Parts for Valves See the Agilent catalog for consumables and supplies for a more complete listing, or visit the Agilent Web site for the latest information (www.agilent.com/chem/supplies). Table 43 Valve supplies Description Part number Valves, gas sampling...
  • Page 267: Exploded Parts View Of Gc Rotary Valves

    17 Maintaining a Valve Exploded Parts View of GC Rotary Valves Exploded Parts View of GC Rotary Valves Valve actuator Upper valve box assembly Valve body Rotor index pin Valve rotor Preload assembly Figure 51. GC rotary valve exploded parts Maintaining Your GC...
  • Page 268: To Replace A Gas Sampling Valve Loop

    17 Maintaining a Valve To Replace a Gas Sampling Valve Loop To Replace a Gas Sampling Valve Loop 1 Gather the following: • Replacement sample loop, see “Consumables and Parts for Valves” on page 266. • 1/4-inch wrench • Vacuum cleaner 2 Place GC in maintenance mode: Maintenance >...
  • Page 269 17 Maintaining a Valve To Replace a Gas Sampling Valve Loop 10 Install the new sample loop. 11 Repressurize the sample loop and check for leaks. 12 Install the upper valve box. See “To Install the Upper Valve Box” on page 275. 13 Exit maintenance mode: Select Maintenance >...
  • Page 270: To Align A Rotary Valve Rotor

    17 Maintaining a Valve To Align a Rotary Valve Rotor To Align a Rotary Valve Rotor 1 Gather the following: • Flathead screwdriver • 3-mm hex key wrench • T-20 Torx screwdriver 2 Set the oven and valve box heated zones to a safe handling temperature (25 °C). 3 Set all valves to Off.
  • Page 271: To Replace A Rotary Valve In The Valve Box

    17 Maintaining a Valve To Replace a Rotary Valve in the Valve Box To Replace a Rotary Valve in the Valve Box Do not install a liquid sampling valve (LSV) in the valve box if you plan to heat the box above WAR N IN G 75 °C.
  • Page 272 17 Maintaining a Valve To Replace a Rotary Valve in the Valve Box 11 Use needle-nosed pliers to move the valve rotor index pin of the valve counterclockwise until the pin touches the valve stop Off position. Valve rotor index pin Valve stop Off position 12 Plumb the new valve using the existing fittings.
  • Page 273 17 Maintaining a Valve To Replace a Rotary Valve in the Valve Box 16 Perform an inlet Leak & Restriction test and reset the maintenance counters. 17 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance > Maintenance Mode > Finished. 18 Restore the analytical method.
  • Page 274: To Remove The Upper Valve Box

    17 Maintaining a Valve To Remove the Upper Valve Box To Remove the Upper Valve Box 1 Gather a T-20 Torx screwdriver. 2 Set the valve box to a safe handling temperature (25 °C), or prepare for maintenance: Maintenance > Instrument > Perform Maintenance > Maintenance Mode > Start Maintenance.
  • Page 275: To Install The Upper Valve Box

    17 Maintaining a Valve To Install the Upper Valve Box To Install the Upper Valve Box 1 Gather the following: • T-20 Torx screwdriver • 3-mm hex key wrench • Flathead screwdriver 2 Verify that all valve rotors are in the full counterclockwise position (valve Off). 3 For each actuator that mates with a newly installed valve: a Loosen the adjustment set screw.
  • Page 276 17 Maintaining a Valve To Install the Upper Valve Box Maintaining Your GC...
  • Page 277: A Swagelok Connections

    Swagelok Connections Making Swagelok Connections 278 Using a Swagelok Tee 281 The gas supply tubing is attached with Swagelok fittings. If you are not familiar with Swagelok connections, review the following procedures. Maintaining Your GC...
  • Page 278: Making Swagelok Connections

    Swagelok Connections Making Swagelok Connections Making Swagelok Connections Objective To make a tubing connection that does not leak and that can be taken apart without damaging the fitting. Materials needed: • 1/8-inch (or 1/4-inch, if used) preconditioned copper tubing • 1/8-inch (or 1/4-inch, if used) Swagelok nuts •...
  • Page 279 Swagelok Connections Making Swagelok Connections 5 Push the tube fully into the plug, then withdraw it approximately 1 to 2 mm as shown in Figure Front ferrule Insert tubing fully Back ferrule Withdraw tubing 1 to 2 mm Tighten nut Figure 54.
  • Page 280 Swagelok Connections Making Swagelok Connections 9 Remove the plug from the fitting. To connect the tubing, with nut and ferrules, to another fitting, finger-tighten the nut, then use a wrench to tighten it 3/4 (1/8-inch fittings) or 1-1/4 (1/4-inch fittings) of a turn. 10 Both correctly- and incorrectly-swaged connections are shown in Figure 57.
  • Page 281: Using A Swagelok Tee

    Swagelok Connections Using a Swagelok Tee Using a Swagelok Tee To supply gas from a single source to more than one input, use a Swagelok Tee. Do not combine valve actuator air with flame ionization air. The valve action will cause major NOTE upsets in the detector signal.
  • Page 282 © Agilent Technologies, Inc. 2019 First edition, January 2019 *G2790-90015* G3540-90015...

Table of Contents