A Accessories and platepads
Setting up a Trash or Filter Plate Holder accessory
Configuring the Filter Plate Holder in the software
IMPORTANT
136
To configure the Filter Plate Holder:
1
In Bravo Diagnostics, click the Profiles tab, and verify that the correct profile is
initialized.
2
Click the Configuration tab.
3
In the Locations list, select the deck location of the installed accessory.
Only deck location 4 or 6 are options for this accessory.
4
In the Location is configured as list, select Accessory. The Accessories Wizard
appears.
5
In the Location list, confirm the deck location of the accessory, and then click Next.
6
In the Accessory list, select Filter Plate Holder, and then click Next.
7
Click Finish.
8
When a message appears and asks whether to move the teachpoint to a safe Z
height, do one of the following:
Click No if you verified the accuracy of the platepad teachpoint before installing
and configuring the accessory, no further adjustment to the teachpoint should
be required.
Click Yes if you plan to reteach this location.
The image of the accessory appears at the specified deck location in the
Configuration tab, as the following figure shows.