Related information
Planning for the protocol run
Reviewing the protocol
Bravo Platform User Guide
To recover from an emergency stop:
1
Turn the red button on the emergency-stop pendant clockwise to release it.
2
In the VWorks message box, click one of the action buttons to re-enable the
motors in the head.
If you want to...
Abort the command or task
Attempt re-execution of the command or task
Ignore the current command or task and proceed to the next
command or task
For information about...
Pausing and continuing a run
Safety hazards
Reporting problems
Before you start a run, make sure you review the protocol and determine the
following:
•
The correct liquid-handling head is installed.
To avoid variability problems, use only Agilent Technologies brand tips.
•
The accessories and labware required for the protocol are where they should be
positioned.
•
The reservoirs that must be filled.
3 Preparing for a protocol run
Planning for the protocol run
Then click...
Abort
Retry
Ignore
See...
VWorks Automation Control User Guide
•
Automation Solutions Products
General Safety Guide
•
Safety and installation guide for your
model Bravo Platform, for example,
G5562A, G5563A Bravo Platform
Safety and Installation Guide
"Reporting problems" on page 87
57