To configure the Vacuum Filtration Station:
1
In Bravo Diagnostics, click the Profiles tab, and verify that the correct profile is
initialized.
2
Click the Configuration tab.
3
In the Locations list, select the deck location of the Vacuum Filtration Station base.
4
In the Location is configured as list, select Accessory. The Accessories Wizard
opens.
5
In the Location list, confirm the deck location of the Vacuum Filtration Station base,
and then click Next.
6
In the Accessory list, select Vacuum Filtration Station, and then click Next.
7
In the "Vacuum Filtration Station" properties table, set the following:
a
Serial Port. The com port on the controlling computer to which the
communications cable from the pinch-valve module or the pump connects.
b
Select one of the Configuration options:
Always assembled. The station consists only of the base and the collar,
or the robot will not assemble or disassemble the station components
during a protocol run. If you select Always assembled, select the Collar
type. The options are Standard (38 mm) or Deep (67.5 mm).
Start disassembled. The station components are at different deck
locations and you want to use the Assemble Vacuum and Disassemble
Vacuum tasks during a protocol run.
c
Pump connected. Choose one of the following:
Pump
Description
Standard
Select this option if the station is connected to the pinch-
valve module.
ME4CVario
Select this option if the station is connected to the ME4C
NT VARIO Vacuum Pump.
If you select ME4CVario, set the Pump Pressure Units to match
the unit setting on the VARIO pump controller. The options are
mbar, Torr, and hPa.