B Autofilling accessories
Setting up a Weigh Station
IMPORTANT
Related information
184
b
In the Accessory Diagnostics dialog box, click Set Range.
The digital display shows a sensor reading that corresponds to the current
weight of the reservoir that is on the Weigh Station.
To avoid overfilling the reservoir, ensure that the full range is set while the fluid level is
below the top of the reservoir.
8
Verify that the Level reading decreases as the liquid level decreases while you run
the designated empty pump to remove any liquid from the reservoir.
9
Click OK.
10
On the Profiles tab, click Update this profile.
For information about...
Starting up and shutting down the
Bravo Platform
Opening Diagnostics
Installing the Pump Module
Configuring the autofilling
function
Editing teachpoints
Using an accessory in a protocol
See...
"Starting up and shutting down" on page 17
"Opening Bravo Diagnostics" on page 24
Pump Module User Guide
"Configuring an autofilling station for the Bravo
Platform" on page 177
"Setting teachpoints" on page 34
VWorks Automation Control User Guide
Bravo Platform User Guide