AM Mix task parameters

About this topic
Parameter description
Bravo Platform User Guide
This topic describes the task parameters for the AM Mix task, which is available only
for the Bravo 96AM Head.
The AM Mix task is designed for aspirating and dispensing fluids through AssayMAP
Bravo cartridges that are mounted on the Bravo 96AM Head.
To run the AM Mix task from Diagnostics, see
Diagnostics" on page
100. To use the task in a protocol, see the
Control User
Guide.
The AM Mix task has the following parameters.
Parameter
Volume (µL)
Pre-aspirate volume (µL)
Blowout volume (µL)
Liquid class
Override flow rates from
liquid class
E Diagnostics quick reference

AM Mix task parameters

"Performing a task using Bravo
Description
The volume of liquid to be mixed in each well.
The volume of air to be drawn before the pipette
tips enter the liquid.
The volume of air to dispense after the main
volume has been dispensed while the tips are still
in the wells.
Typically, the blowout volume is the same as the
pre-aspirate volume.
The pipetting velocity and accuracy.
IMPORTANT
To ensure consistent pipetting,
always select a liquid class for liquid-handling
tasks.
The option to override the aspirate and dispense
velocity in the selected liquid class. Selecting this
option enables you to specify a value for the flow
rate without changing the liquid class.
VWorks Automation
281