E Diagnostics quick reference

Pin Tool task parameters

Pin Tool task parameters

About this topic
Parameter descriptions
268
This topic describes the task parameters for the Pin Tool task.
To run the Pin Tool task, see
page
100.
The Pin Tool task contains the following parameters.
Parameter
Dwell time (s)
Descriptive label
Liquid class
Pipette technique
"Performing a task using Bravo Diagnostics" on
Description
The time duration that the pins remain at the
specified height (First distance or Second
distance) within the well.
For example, you might start with the following
values:
Adsorb, Dispense into fluid, or Mix—0.5 s or
longer for more viscous fluids
Blot—2 s, or longer for more viscous fluids
A text label that you can add to the task icon in the
protocol. Click the arrow in the Descriptive label
box to choose an option.
The options include:
Enter a JavaScript variable or script.
Use a predefined label: Adsorb, Dispense,
Wash, Blot, or Mix
Type your own label in the box.
A parameter that you can use to control the
accuracy and the speed of the pin tool as it moves
into and out of the wells.
IMPORTANT
To ensure consistent pipetting,
always select a liquid class for liquid-handling
tasks.
The pipette location offset you want to use for the
Pin Tool task.
The list of pipette techniques are defined in the
Pipette Technique Editor.