Agilent Technologies Bravo Platform User Manual page 169

Hide thumbs Also See for Bravo Platform:
Table of Contents

Advertisement

A Accessories and platepads
Installing a Heating Shaking Station
158
4
Place the Heating Shaking Station on top of the adapter pad.
Ensure that the station sits level on the adapter pad and the safety label
the front of the workspace.
5
Insert the two M3-x-18-mm socket-head cap screws into the holes at the front and
back end of the adapter pad and use the 2.5-mm hex wrench to tighten the screws.
6
Determine if the pairs of alignment tabs (figure, item 1) in each corner of the plate
nest require adjustment as follows:
a
Position a labware in the plate nest, ensuring that it is fully seated.
b
Make sure that the labware fits in the nest so that it is not too loose or too tight.
To test the fit, lift the labware and re-seat it several times.
c
If necessary, use a flathead driver to adjust the setscrews so that the tabs are
recessed or extended, as needed.
7
Connect the cables:
a
Connect the cable from the Heating Shaking Station to the Inheco STC or MTC
Controller.
Note: The accessory cable is color-coded to match the card installed in the
STC or MTC.
b
Connect the STC or MTC Controller power and communication cables. See the
Inheco user documentation for details on how to connect the STC or MTC.
c
Press the STC or MTC Controller power switch to the on (|) position.