Mix task parameters

About this topic
Parameter descriptions
Bravo Platform User Guide
For information about...
Editing the labware classes, liquid
library, or pipetting techniques
Performing a diagnostic task
Reporting a problem
The Mix task aspirates and dispenses liquid multiple times to mix it. You can specify
different well-bottom distances for the aspirate and dispense actions. This topic
describes the task parameters.
To run either task, see
"Performing a task using Bravo Diagnostics" on page
The Mix task parameters include the following.
Parameter
Volume
(0–200 µL)
Pre-aspirate volume
(0–200 µL)
Blowout volume
(0–200 µL)
Liquid class
Mix cycles ((0–100)
E Diagnostics quick reference
See...
VWorks Automation Control Setup Guide
"Performing a task using Bravo Diagnostics"
on page 100
"Reporting problems" on page 87
Description
The volume of liquid to be mixed in each well.
The volume of air to be drawn before the pipette
tips enter the liquid.
Specifies the volume of air to dispense after the
main volume has been dispensed while the tips
are still in the wells.
Typically, the blowout volume is the same as the
pre-aspirate volume.
The pipetting speed and accuracy.
IMPORTANT
To ensure consistent pipetting,
always select a liquid class for liquid-handling
tasks.
The number of times to repeat the aspirate-and-
dispense cycle.

Mix task parameters