4 Maintenance and troubleshooting
Replacing the fuse
Procedure
74
To replace the fuse in the power switch:
1
Shut down the Bravo Platform, and unplug the power cable from the rear panel
connector.
2
At the rear panel power switch enclosure, use a small flat-head screwdriver (2.5
mm) to pry open the tab on the fuse enclosure and open the enclosure cover.
00220
Bravo
Back
3
Insert the screwdriver head in the enclosure notch to dislodge the fuse cartridge.
Slide the fuse cartridge all the way out of the enclosure.
4
Replace the fuse in the cartridge.
5
Slide the fuse cartridge back into the fuse enclosure.
6
Press the enclosure cover securely into the closed position.
7
Plug in the power cable at the rear panel connector, and then start up the Bravo
Platform.
Bravo Platform User Guide