A Accessories and platepads
Setting up the Thermal Station (cooling pad)
Completing the hardware setup
154
a
Tighten the setscrew on the end tab.
b
Tighten the setscrews in the two side tabs.
7
Install the side brackets as follows:
a
Position the brackets, aligning the slots in the brackets with the screw holes in
the adapter pad.
b
Install the two hex screws in each bracket, and then apply downward pressure
on the bracket tops while tightening the hex screws.
Ensure that both brackets are holding the thermal pad securely in place.
After installing the Thermal Station, the following setup tasks must be completed
before the accessory is ready for use.
Step
Procedure
1
Connect one end of the marprene tubing to the
cooling unit drain and place the other end in a
waste container.
2
Install the straight ends of the thermal tubing
into the cooling pad, and connect the other ends
into the cooling unit.
3
Connect the power cord from cooling cube to an
AC outlet.
4
Fill the liquid chilling unit as follows:
a
Turn on the unit.
b
Watch for the Liquid Level Low message
to display on the front panel.
c
Turn off the unit and fill it with 25%
EtOH or cooling fluid recommended by
manufacturer.
d
Repeat steps a to c. The unit should
hold 600 to 700 mL.
See...
Liquid chilling unit
manufacturer's user
documentation
Liquid chilling unit
manufacturer's user
documentation
Liquid chilling unit
manufacturer's user
documentation
Liquid chilling unit
manufacturer's user
documentation
Bravo Platform User Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents