Pump Reagent Parameters - Agilent Technologies Bravo Platform User Manual

Hide thumbs Also See for Bravo Platform:
Table of Contents

Advertisement

E Diagnostics quick reference

Pump Reagent parameters

Pump Reagent parameters

About this topic
Parameter descriptions
258
The Pump Reagent task fills or empties a reservoir or wash tray by pumping for a
specified number of seconds. If the reservoir is on a Weigh Station, the pump stops
fluid flow when the target weight is reached. This topic describes the task
parameters.
To run the Pump Reagent task, see
page
100.
The Pump Reagent task has the following parameters.
Parameter
Reservoir mode
Pump speed
Pump on time
Use weigh station/shelf
Weigh station/shelf action
threshold
Weigh station stop action
threshold
"Performing a task using Bravo Diagnostics" on
Description
The action of the task:
Fill
Empty
The speed, in percent of maximum, at which to
pump the reagent.
The duration of time, in seconds, to run the pump:
If the Use weigh station/shelf option is selected,
this value specifies the maximum time (seconds)
to run the pump. If the pump runs for the full
duration without reaching the weigh station/shelf
threshold, an error occurs.
The option to use the Weigh Station or Weigh
Shelf.
The minimum fluid weight, in percent of the full
weight that was calibrated on the Weigh Station
or Weigh Shelf.
For example, you can set the minimum threshold
at 45% so that when the fluid reaches 45% of the
full weight, fluid starts to pump into the reservoir.
The maximum fluid weight, in percent of the full
weight that was calibrated on the Weigh Station
or Weigh Shelf.
For example, you can set the stop threshold at
60% so that when the fluid reaches 60% of the full
weight, fluid starts to drain or pump out of the
reservoir.
Bravo Platform User Guide