A Accessories and platepads
Installing and setting up a Vacuum Filtration Station
132
Figure
Accessory Diagnostics dialog box for the ME4C NT VARIO Vacuum Pump
4
To configure the pump:
a
In the Get/Set the pressure in list, select the desired unit of measure (mbar, Torr,
hPa, mmHg, cmHg, and inHg). If the selected units differ from the
configuration, the software will convert the values.
b
Set the Vent valve to Open.
c
Set the Pump to Off.
d
Type the Target pressure (differential).
e
Click Set baseline pressure.
5
To start the vacuum:
a
Set the Vent valve to Close.
b
Set the Pump to On.
c
Set the Target pressure.
d
Click Execute.
6
To stop the vacuum:
a
Set the Vent valve to Open.
b
Set the Pump to Off.
c
Click Execute.
Bravo Platform User Guide