Step 8. Run the Operator's Check

The operator's check should be performed when you first receive your analyzer,
and any time you wish to have confidence that the analyzer is working
properly. The operator's check does not verify performance to specifications,
but should give you a high degree of confidence that the instrument is
performing properly. The only equipment required is an open or short
calibration standard that mates with the front panel test ports. The open or
short calibration standard is NOT included with the PNA.
If a "Log Error" is displayed during the verification stating that a
'filename.csv' is already open and in use, verify that you only have a single
verification window open. This can be verified by mousing over the
verification icon on the toolbar as the bottom of the display. If there are
two verification windows open, close the window on the right most side,
which is the most recently opened window. And, resume the verification.
— On the A models: Press System, then [Service] softkey, then [Verification]
— On the B models: Press System, then [Service] softtab, then [Verification],
— For assistance, use the Help button displayed on the screen of the

Step 9. Turn Off the Analyzer

To turn the PNA OFF, BRIEFLY press the front panel On/Standby switch. The
PNA will then begin to shutdown
Shutdown
The On/Standby indicator changes from green to yellow when power is Off.
The current instrument state is NOT saved to the PNA solid state drive.
When the PNA is powered ON, a full system boot-up is performed and the PNA
powers-up in the Preset settings.
Repeatedly doing either of the following two actions may cause damage to
the PNA solid state drive.
1. Only in an emergency should the power cord be removed from the PNA
2. If the PNA is locked and you cannot operate the mouse or keypad, turn
Installation Note E8356-90001
softkey, then [Operator's Check]. Follow the prompts.
then [Operator's Check]. Follow the prompts.
Operator's Check.
without first being put in Shutdown mode.
OFF the PNA by pressing and holding the power button for at least four
seconds.