Troubleshooting and Diagnostics

Overview of the Module's Indicators and Test Functions

Overview of the Module's Indicators and Test Functions

Status Indicators
The module is provided with two status indicators which indicate the
operational state (prerun, run, and error states) of the module. The status
indicators provide a quick visual check of the operation of the module.
Error Messages
In the event of an electronic, mechanical or hydraulic failure, the module
generates an error message in the user interface. For each message, a short
description of the failure, a list of probable causes of the problem, and a list of
suggested actions to fix the problem are provided (see chapter Error
Information).
Not-Ready Messages
During the wait for a specific condition to be reached or completed the
detector will generate a not-ready message. For each message a short
description is provided (see
Refractive Index Calibration
Refractive index calibration is recommended after exchange of the optical unit
to ensure correct operation of the detector. The procedure uses a solution of
known refractive index compared to LC grade water (see
Calibration"
90
"Not-Ready Messages"
on page 118).
on page 114).
"Refractive Index