6
Troubleshooting and Diagnostics
Overview of the Detector's Indicators and Test Functions
Status Indicators
The detector is provided with two status indicators which indicate the
operational state (prerun, run, and error states) of the detector. The status
indicators provide a quick visual check of the operation of the detector (see
page
Error Messages
In the event of an electronic, mechanical or hydraulic failure, the detector
generates an error message in the user interface. For each message, a short
description of the failure, a list of probable causes of the problem, and a list of
suggested actions to fix the problem are provided (see
the Service
Test Functions
A series of test functions are available for troubleshooting and operational
verification after exchanging internal components (see
Service
Wavelength Verification / Recalibration
Wavelength recalibration is recommended after exchange of lamps and flow
cells, maintenance of flow cells, repair of internal components, and on a
regular basis to ensure correct operation of the
the deuterium alpha and beta emission lines for wavelength calibration (see
"Wavelength Verification and Recalibration"
Diagnostic Signals
The detector has several signals (internal temperatures) that can be used for
diagnosing baseline problems (see
88
51).
Manual.).
Manual).
detector
on page 138).
"Diagnostic Signals" in the Service
1200 Series DAD and MWD User Manual
"Error Information" in
"Test Functions" in the
detector
. The
uses
Manual).