FAG Detector III User Manual page 51

F'is trendline 3
Hide thumbs Also See for Detector III:
Table of Contents

Advertisement

calibration and want to use the balancing function on a different Detector in the
meantime.
To activate the balancing function on the Detector, start the Trendline software.
Connect the Detector to the PC using the serial cable. Plug the dongle into a free
USB port. Switch on the Detector and select the Trendline menu item Detector >
balancing activation. The following window is now displayed:
The serial number of the connected Detector is displayed at the top. The dongle
that has been identified is displayed at the bottom. The identified version should
start with "3" in the case of Detector III. The serial number of the Detector for
which this dongle has already been used or, if the dongle has not yet been used,
Empty is displayed in the Used for Detector: column. Now select the required
dongle from the Type column (should be displayed as "Empty" in the "Used for
Detector" column). You can now enable the balancing functionality via Activate
balancing for connected Detector. A new line Balancing should now be visible
and the serial number of the Detector is displayed in the dongle line.
Please note the following:
Trendline 3
51