You can accept the suggested name or enter a different name.
Pitch: The pitch at the reference point is needed to display the position of the
measuring point in the amplitude/phase graph. If at least two measuring points
are already available, the Detector automatically suggests a value composed
of the pitch of the last measuring point and the pitch between the two last
measuring points. You can accept the suggestion or enter a different value.
The measurement is started. The Detector measures the rotational speed as
well as amplitude and phase of the vibration. A bar showing mean variance
displays the stability of the values. You can reset the mean by pressing the
function button
to restart mean calculation.
Press the Enter button
time.
After completing the measurement, you can select a comment. In the comment
list, you can create a new comment
Managing comments for measurements
the edit field for the comment input is shown. Confirm your settings with the
Enter button
.
Select Values are OK.
The values for the measuring point are saved along with the comment. You can
now continue or stop the measurement at the next measuring point.
The numb er of measuring points is limited to 255. Once this
numb er is reached, the Detector displays a warning and
automatically stops the measurement.
If there is not enough memory for an additional measuring point,
the Detector also displays a warning and automatically stops the
measurement.
After completing the measurement, the Detector displays all measured values in
a table.