FAG Detector III User Manual page 264

F'is trendline 3
Hide thumbs Also See for Detector III:
Table of Contents

Advertisement

Error message
Description
or sw itch on the
temperature sensor.
Temperature too
The measured
high
temperature is bey ond
the technical limit of the
temperature sensor.
No memory left
Cannot create any more
for sav ing data
measurement, balancing
step or new data block
for run up/coast dow n
because memory is full.
Warning: battery
The battery is nearly
low
empty .
The battery is
The battery is empty ,
empty Please
the Detector cannot be
charge
sw itched on.
No memory
No memory has been
allocated
allocated for
configurations that hav e
been sent to the
Detector.
No template
No balancing, CM or run
loaded
up/coast dow n template
has been sav ed.
No configuration
No balancing, CM or run
loaded!
up/coast dow n
configuration has been
sav ed.
No free
No free measurements
measurements
hav e been performed
sav ed
y et.
No balancing
The Detector is
configuration
attempting to sav e time
stored.
signals but cannot
because of insufficient
memory .
No memory
The Detector is
av ailable for time
attempting to sav e time
264
Cause of the fault / solution
Temperature measurement range: -15 °C to +550 °C
Transfer the data to Trendline and transfer the last balancing
configuration or route back to the Detector.
Please recharge
the accumulator before continuing to
218
w ork w ith the Detector. Otherw ise, the Detector could shut
dow n w ithout any w arning so as to protect the accumulator.
The battery must be recharged
The Trendline database may contain an error. Please
contact support@fis-serv ices.com.
You are try ing to perform a free measurement, but there is
no template is stored in the Detector. Templates are only
attached if y ou click in Trendline on Detector > Send route
.
You w ish to carry out balancing, a CM measurement or a
run up/coast dow n procedure but hav e not y et transferred a
configuration to the Detector. In Trendline click on Detector
> Send configuration or Detector > Send route.
You are attempting to v iew free balancing, CM
measurements or run up/coast dow n measurements on the
Detector although neither of these ty pes of measurement has
been carried out.
Send a balancing configuration to the Detector (click in
Trendline on Detector > Send configuration or
Detector > Send route) before y ou select a configuration
from the balancing menu of the Detector.
This can occur if y ou w ant to sav e only those time signals
occurring in the ev ent of an alarm. It is also possible that
218
.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents