Delete Measured Data - FAG Detector III User Manual

F'is trendline 3
Hide thumbs Also See for Detector III:
Table of Contents

Advertisement

ICP sensor test
The ICP sensor test checks the following cases based on the bias voltage of the
sensor connected to the BNC1 connector:
The voltage is within the specified range: The sensor is functional.
The voltage is greater than maximum sensor voltage: The sensor cable is
defective or no sensor is connected.
The voltage is between 0 and minimum sensor voltage: The sensor is
defective.
After starting the ICP sensor test, it may take a few seconds, until all
transient effects are concluded in the sensor.

6.13 Delete measured data

You can delete the measuring data of a CM, balancing, run up/coast down or
amplitude/phase measurement on the Detector. Therefore select the desired
level (hall, machine or measuring point) or mark directly a measurement to
delete.
Please note, that the measurement data will b e irrevocab ly lost if
you confirm the deleting process. You cannot recover the data!
Delete measurement data on the Detector
Select the level below which the measured data should be deleted.
Press the function button .
Select Delete data and
confirm the deleting process with the Enter button
The following table describes the deleting of CM measurement data on each
level. The deleting procedure is the same for balancing, run up/coast down or
amplitude/phase measurements.
Detector III
.
261